A Pilot Study on Parameter Setting of VisiTag™ Module during Pulmonary Vein Isolation
المؤلفون المشاركون
Wang, Yu-Chuan
Huang, Bo
Li, Kang
He, Peng-Kang
Chen, Er-Dong
Xia, Yu-Long
Jiang, Jie
Sheng, Qin-Hui
Zhou, Jing
Ding, Yan-Sheng
المصدر
Cardiology Research and Practice
العدد
المجلد 2018، العدد 2018 (31 ديسمبر/كانون الأول 2018)، ص ص. 1-5، 5ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2018-10-29
دولة النشر
مصر
عدد الصفحات
5
التخصصات الرئيسية
الملخص EN
Objectives.
To identify optimal predefined criteria (OPC) for filters of the VisiTag™ module in the CARTO 3 system during pulmonary vein isolation (PVI).
Methods.
Thirty patients with atrial fibrillation (AF) who experienced PVI first were enrolled.
PVI was accomplished by using a Thermocool SmartTouch catheter.
Ablation lesions were tagged automatically as soon as predefined criteria of the VisiTag™ module were met.
OPC should be that ablation with the setting resulting in the conduction gap (CG) as few as possible, while contiguous encircling ablation line (CEAL) without the tag gap (TG) on the 3D anatomic model as much as possible.
Result(s).
When ablation with parameter setting is being catheter movement with a 3 mm distance limit for at least 20 s and force over time (FOT) being off, there were 60 CEAL without TG on the 3D anatomic model.
However, 26 CGs were found.
After changing FOT setting to be a minimal force of 5 g with 50% stability time, 22 TGs were displayed.
Of them, 20 TGs were accompanied by CGs.
On reablation at sites of TG with changed parameter setting, 18 CGs were eliminated when 20 TGs disappeared.
When reablation with FOT is being a minimal force of 10 g with 50% stability time, 6 remaining CGs were eliminated.
However, there was no CEAL.
With a mean of follow-up 10.93 months, 2 patients with persistent AF suffered AF recurrence.
Conclusion.
A 3 mm distance limit for at least 20 s and FOT being a minimal force of 5 g with 50% stability time might be OPC for the VisiTag™ module.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Wang, Yu-Chuan& Huang, Bo& Li, Kang& He, Peng-Kang& Chen, Er-Dong& Xia, Yu-Long…[et al.]. 2018. A Pilot Study on Parameter Setting of VisiTag™ Module during Pulmonary Vein Isolation. Cardiology Research and Practice،Vol. 2018, no. 2018, pp.1-5.
https://search.emarefa.net/detail/BIM-1152118
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Wang, Yu-Chuan…[et al.]. A Pilot Study on Parameter Setting of VisiTag™ Module during Pulmonary Vein Isolation. Cardiology Research and Practice No. 2018 (2018), pp.1-5.
https://search.emarefa.net/detail/BIM-1152118
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Wang, Yu-Chuan& Huang, Bo& Li, Kang& He, Peng-Kang& Chen, Er-Dong& Xia, Yu-Long…[et al.]. A Pilot Study on Parameter Setting of VisiTag™ Module during Pulmonary Vein Isolation. Cardiology Research and Practice. 2018. Vol. 2018, no. 2018, pp.1-5.
https://search.emarefa.net/detail/BIM-1152118
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-1152118
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر