Reproducibility of BiliCare™ Transcutaneus Bilirrubin Meter in Mexican Newborns

المؤلفون المشاركون

Oldak, David
García, Gabriela
Gonzalez, Estela E.
Aillon, Enver
Falcon, Juan C.
Ayala, Ernesto
Oldak, Bernardo

المصدر

International Journal of Pediatrics

العدد

المجلد 2019، العدد 2019 (31 ديسمبر/كانون الأول 2019)، ص ص. 1-5، 5ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2019-01-01

دولة النشر

مصر

عدد الصفحات

5

التخصصات الرئيسية

الطب البشري

الملخص EN

Background.

Newborn hyperbilirubinemia is considered a worldwide health problem that demands medical evaluation.

Noninvasive transcutaneous bilirubin (TcB) has been used as a screening method with different devices but there has not been any evaluation of reproducibility of the same brand devices.

The BiliCare™ system is evaluated to demonstrate consistency between measurements with four different devices.

Methods.

336 TcB measurements were obtained with four BiliCare™ devices in 21 Mexican icteric newborns with a mean postnatal age of 44.1 hours of life and 38 weeks of gestation (33–41).

Two measurements were taken in the same ear alternatively at the scaphoid fossa with each device.

TcB values were compared between devices.

Validity was compared with total serum bilirubin (TB).

Results.

intraclass correlation coefficient demonstrates a minimum limit in the study of 0.945 and maximum of 0.988 with the same device.

Correlations with serum and between devices gave results above 0.932.

Conclusions.

BiliCare™ transcutaneous bilirubin measurement instrument has very good intra- and interdevice reproducibility; also correlation of TcS with serum bilirubin gave statistically the same results.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Oldak, David& García, Gabriela& Gonzalez, Estela E.& Aillon, Enver& Falcon, Juan C.& Ayala, Ernesto…[et al.]. 2019. Reproducibility of BiliCare™ Transcutaneus Bilirrubin Meter in Mexican Newborns. International Journal of Pediatrics،Vol. 2019, no. 2019, pp.1-5.
https://search.emarefa.net/detail/BIM-1167571

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Oldak, David…[et al.]. Reproducibility of BiliCare™ Transcutaneus Bilirrubin Meter in Mexican Newborns. International Journal of Pediatrics No. 2019 (2019), pp.1-5.
https://search.emarefa.net/detail/BIM-1167571

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Oldak, David& García, Gabriela& Gonzalez, Estela E.& Aillon, Enver& Falcon, Juan C.& Ayala, Ernesto…[et al.]. Reproducibility of BiliCare™ Transcutaneus Bilirrubin Meter in Mexican Newborns. International Journal of Pediatrics. 2019. Vol. 2019, no. 2019, pp.1-5.
https://search.emarefa.net/detail/BIM-1167571

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1167571