This study was aimed at analyzing data and evaluating the accuracy of a new subcutaneous continuous glucose-monitoring device by referring to finger-pricking measurement.

The data were obtained from 7 diabetic patients.

An improved implanted flex sensor was used to measure the interstitial glucose concentration every 3 min within 6 days, and five finger-pricking samples were collected every day for comparison.

A periodic glucose change happened every day.

2.45% of CGM values were in the hypoglycemic range (<70 mg/dl), 55.4% were in the normal range (70-180 mg/dl), and 42.15% were in the hyperglycemic range (>180 mg/dl).

The interstitial glucose concentrations (n=204) were well linearly correlated with the capillary glucose concentrations (r=0.94, P<0.01), but a delay occurred between 10 and 40 minutes within two measurements (CGM later), and the individual average delay times were relatively close to 24.6 minutes.

Clarke’s error grid analysis showed that 86.7% of the points fell in zone A, 12.7% fell in zone B, and only 0.6% fell in zone D.

An overall MARD was 10.22%.

This study demonstrated that the CGM was accurate and highly reliable; thus, it constituted a new device for continuous glucose monitoring in diabetic patients.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Cai, Lijun& Ge, Wancheng& Zhu, Zhigang& Zhao, Xueling& Li, Zhanhong. 2019. Data Analysis and Accuracy Evaluation of a Continuous Glucose-Monitoring Device. Journal of Sensors،Vol. 2019, no. 2019, pp.1-8.
https://search.emarefa.net/detail/BIM-1191232

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Cai, Lijun…[et al.]. Data Analysis and Accuracy Evaluation of a Continuous Glucose-Monitoring Device. Journal of Sensors No. 2019 (2019), pp.1-8.
https://search.emarefa.net/detail/BIM-1191232

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Cai, Lijun& Ge, Wancheng& Zhu, Zhigang& Zhao, Xueling& Li, Zhanhong. Data Analysis and Accuracy Evaluation of a Continuous Glucose-Monitoring Device. Journal of Sensors. 2019. Vol. 2019, no. 2019, pp.1-8.
https://search.emarefa.net/detail/BIM-1191232

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1191232