Are tubes containing sodium fluoride still needed for the measurement of blood glucose in hospital laboratory practice ?
المؤلفون المشاركون
al-Kindi, Manal
Mula Abid, Wad Allah
al-Kharusi, Amirah
al-Lawati, Nuha
المصدر
العدد
المجلد 29، العدد 6 (31 ديسمبر/كانون الأول 2014)، ص ص. 404-407، 4ص.
الناشر
المجلس العماني للاختصاصات الطبية
تاريخ النشر
2014-12-31
دولة النشر
سلطنة عمان
عدد الصفحات
4
التخصصات الرئيسية
الملخص EN
Objectives: To compare glucose values obtained using sodium flouride (NaF) tubes and serum separation tubes (SST) in a tertiary care hospital laboratory setting.
Methods: This study was conducted at the Clinical Biochemistry Laboratory, Royal Hospital, Oman.
During the study period (1 September–30 November 2013), 50 pairs (one NaF tube and one SST) of patient’s blood specimens were randomly collected.
Following separation of plasma (NaF tubes) and serum (SST), glucose concentrations were measured by hexokinase assay using the Architect c8000.
Fifteen pairs of these tubes were kept in the refrigerator at 4°C and plasma/serum glucose concentrations were measured daily up to seven days after collection.
Results: Comparing plasma (NaF) and serum (SST) results of glucose values (n=50) showed an average difference of 0.00mmol/L (range -0.60 to +0.60mmol/L).
Bland Altman analysis gave a non-significant constant bias of 0.10 ±0.195mmol/L (bias ±SD).
Pearson correlation between plasma (NaF) and serum (SST) glucose concentrations revealed a significant correlation approaching unity with r2= 0.9991.
No significant differences in glucose values were noted for both plasma and serum in 15 pairs of NaF and SST tubes when analyzed seven days following refrigeration.
Hemolysis was observed in five (10%) NaF tubes compared with two (4%) SST.
Conclusion: There is no difference in glucose values collected from plasma NaF tubes or serum SST, and so SST can be used in hospital laboratory settings as there are practical advantages, including cost-effectiveness and reduction in blood volume drawn when utilizing these tubes for glucose and other tests from a single blood collection tube
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
al-Kharusi, Amirah& al-Lawati, Nuha& al-Kindi, Manal& Mula Abid, Wad Allah. 2014. Are tubes containing sodium fluoride still needed for the measurement of blood glucose in hospital laboratory practice ?. Oman Medical Journal،Vol. 29, no. 6, pp.404-407.
https://search.emarefa.net/detail/BIM-831860
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
al-Kharusi, Amirah…[et al.]. Are tubes containing sodium fluoride still needed for the measurement of blood glucose in hospital laboratory practice ?. Oman Medical Journal Vol. 29, no. 6 (2014), pp.404-407.
https://search.emarefa.net/detail/BIM-831860
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
al-Kharusi, Amirah& al-Lawati, Nuha& al-Kindi, Manal& Mula Abid, Wad Allah. Are tubes containing sodium fluoride still needed for the measurement of blood glucose in hospital laboratory practice ?. Oman Medical Journal. 2014. Vol. 29, no. 6, pp.404-407.
https://search.emarefa.net/detail/BIM-831860
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references : p. 407
رقم السجل
BIM-831860
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر