Determination of oxonium ion using laser diode fluorimeter : (blue purple (405 nm)‎-green (532nm)‎ using photodiode at 900 as a detector and I -IO3 -H3O+ system for quenching of fluorescence

Other Title(s)

تقدير أيون الإكسينيوم باستخدام محلل الحقن الجرياني فلور وميتر بثنائيات وصلة ليزرية : (أزرق-بنفسجي 405 نانومتر)‎-الأخضر (532 نانومتر)‎ باستخدام ثنائي وصلة فوتوني بزاوية 90-IO3- H3 O+ I- كمتحسس لنظام إخماد الفلورسين

Joint Authors

al-Bana, Marwah Abd al-Rida Kazim
al-Awwadi, Nagham Shakir Turki

Source

Iraqi Journal of Science

Issue

Vol. 56, Issue 3B (30 Sep. 2015), pp.2126-2148, 23 p.

Publisher

University of Baghdad College of Science

Publication Date

2015-09-30

Country of Publication

Iraq

No. of Pages

23

Main Subjects

Chemistry

Abstract AR

طورت طريقة جديدة تميزت بالحساسية و السرعة و الدقة لتقدير ايون الاوكسونيوم بواسطة استخدام محلل الحقن الجرياني-فلوروميتير بثنائيات وصلة ليزري.

استندت الطريقة على تحرير اليود مسببا إخماد ضوء الفلورة (الفلورة المستمرة من جزيئة الفلورسين الحامضية عند التشعيع بلمصدر الليزري بطول موجي 405 نانومتر).

تم دراسة المتغيرات الكيميائية و الفيزيائية.

حيث تم اختيار سرعة جريان مثلى 1.3 ملليتر / دقيقة باستخدام محلول الفلورسين كتيار ناقل و حجم انموذج 35 مايكروليتر لصمام الحقن الأول و 39 مايكروليتر لصمام الحقن الثاني.

أظهرت نتائج معالجة البيانات أن مدى الخطية لحامض الهيدروكلوريك، الكبريتيك، البريكلوريك و الترتاريك هي 0.002-0.1 مللي مول لتر-1، 0.01-0.2 ملي مول لتر-1، 0.002-0.5 مللي مول لتر-1 و 0.002-0.7 مللي مول لتر-1 على التوالي بمعامل ارتباط ® = 0.9562، 0.9976، 0.9755 و 0.9579 على التوالي.

حد الكشف 1.28 نغم / انموذج، 17.16 نغم / انموذج، 3.52 نغم / انموذج و 5.25 نغم/ انموذج من التخفيف التدريجي لأقل تركيز في منحني المعايرة.

الانحراف القياسي النسبي لتركيز 0.01 مللي مول لتر-1 أقل من 1 % بتكرارية لثمان مرات.

أجريت المقارنة بين الطريقة المستحدثة و الطريقة التقليدية (Ph-meter) باستخدام الإضافات القياسية بوساطة إجراء اختبار t المزدوج بمسارين المسار الأول على أساس الاختلاف بين نوع الطريقة المستخدمة في التحليل و المسار الثاني استنادا على اختلاف مصدر الحامض و لوحظ من النتائج انه لا يوجد فرق جوهري بين الطريقتين و بين مصدر الحامض المستخدم عند مستوى قناعة 95 %.

Abstract EN

A newly developed sensitive, fast and accurate analytical method characterized by a sensitive, fast and accuracy for the determination of oxonium ion by laser diode fluorimeter-flow injection analysis.

The method is based on the libration of free iodine causing to quench the fluorescence light (continuous fluorescence from fluorescien molecule (free acid)) when irradiated by laser source at 405nm.

Chemical and physical parameters were studied and optimized.

The optimum parameters were 1.3mL/min using fluorescien (free acid) as a carrier stream, 35μL sample volume at valve no.1 and 39μL at valve no.2 and opened valves sequentially.

Data treatment shows that linear range 0.002-0.1mMol.L-1,0.01-0.2mMol.L-1,0.002-0.5mMol.L-1,0.002-0.7mMol.L-1 for HCl, H2SO4,HClO4 and tartaric acid respectively, with correlation coefficient r= 0.9562, 0.9976.

0.9755 and 0.9579.

The limit of detection (S/N=3) 1.28ng/sample, 17.16ng/sample, 3.52ng/sample and 5.25ng/sample for HCl, H2SO4,HClO4 and tartaric acid respectively from the stepwise dilution for minimum concentration of lowest concentration in linear dynamic range of the calibration gragh.

RSD% for the repeatability (n=8) was ˂1% for the determination of HCl, H2SO4,HClO4 and tartaric acid with concentration 0.01mMol.L-1 .

The comparison was made between the newly developed method analysis with the classical method (pH-meter) using the standard addition method via the use of paired t-test was studied at two different paths .First path a comparison between two methods of analysis and second path between the two different acids.

It was noticed that there was no significant difference between different acids and different methods at 95% confidence interval level.

American Psychological Association (APA)

al-Awwadi, Nagham Shakir Turki& al-Bana, Marwah Abd al-Rida Kazim. 2015. Determination of oxonium ion using laser diode fluorimeter : (blue purple (405 nm)-green (532nm) using photodiode at 900 as a detector and I -IO3 -H3O+ system for quenching of fluorescence. Iraqi Journal of Science،Vol. 56, no. 3B, pp.2126-2148.
https://search.emarefa.net/detail/BIM-645117

Modern Language Association (MLA)

al-Awwadi, Nagham Shakir Turki& al-Bana, Marwah Abd al-Rida Kazim. Determination of oxonium ion using laser diode fluorimeter : (blue purple (405 nm)-green (532nm) using photodiode at 900 as a detector and I -IO3 -H3O+ system for quenching of fluorescence. Iraqi Journal of Science Vol. 56, no. 3B (2015), pp.2126-2148.
https://search.emarefa.net/detail/BIM-645117

American Medical Association (AMA)

al-Awwadi, Nagham Shakir Turki& al-Bana, Marwah Abd al-Rida Kazim. Determination of oxonium ion using laser diode fluorimeter : (blue purple (405 nm)-green (532nm) using photodiode at 900 as a detector and I -IO3 -H3O+ system for quenching of fluorescence. Iraqi Journal of Science. 2015. Vol. 56, no. 3B, pp.2126-2148.
https://search.emarefa.net/detail/BIM-645117

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references : p. 2147-2148

Record ID