Assessment of long distance chasing photometer (NAG-ADF-300-2) by estimating the drug atenolol with ammonium molybdate via continuous flow injection analysis
Other Title(s)
تقييم مطياف المسافة التتبعية الطويل من خلال تقدير عقار الأتينولول مع مولبيدات الأمونيوم بوساطة تقنية الحقن الجرياني المستمر
Joint Authors
Turki, Nagham Shakir
Mazal, Ilham Naghimish
Source
Issue
Vol. 17, Issue 1 (31 Mar. 2020), pp.78-92, 15 p.
Publisher
University of Baghdad College of Science for Women
Publication Date
2020-03-31
Country of Publication
Iraq
No. of Pages
15
Main Subjects
Topics
Abstract EN
Atenolol was used with ammonium molybdate to prove the efficiency, reliability and repeatability of the long distance chasing photometer (NAG-ADF-300-2) using continuous flow injection analysis.
The method is based on reaction between atenolol and ammonium molybdate in an aqueous medium to obtain a dark brown precipitate.
Optimum parameters was studied to increase the sensitivity for developed method.
A linear range for calibration graph was 0.1-3.5 mmol/L for cell A and 0.3-3.5 mmol/L for cell B, and LOD 133.1680 ng/100 µL and 532.6720 ng/100 µL for cell A and cell B respectively with correlation coefficient (r) 0.9910 for cell A and 0.9901 for cell B, RSD% was lower than 1%, (n=8) for the determination of atenolol at concentration (0.5, 0.7 and 5) mmol/L respectively.
The results were compared with classical method UV-Spectrophotometric at λ max=270 nm using the standard addition method via the use of t-test, at 95% confidence level.
The comparison of data explain that long distance chasing photometer (NAG-ADF-300-2) is the choice with excellent extended detection and wide application.
American Psychological Association (APA)
Turki, Nagham Shakir& Mazal, Ilham Naghimish. 2020. Assessment of long distance chasing photometer (NAG-ADF-300-2) by estimating the drug atenolol with ammonium molybdate via continuous flow injection analysis. Baghdad Science Journal،Vol. 17, no. 1, pp.78-92.
https://search.emarefa.net/detail/BIM-935378
Modern Language Association (MLA)
Turki, Nagham Shakir& Mazal, Ilham Naghimish. Assessment of long distance chasing photometer (NAG-ADF-300-2) by estimating the drug atenolol with ammonium molybdate via continuous flow injection analysis. Baghdad Science Journal Vol. 17, no. 1 (2020), pp.78-92.
https://search.emarefa.net/detail/BIM-935378
American Medical Association (AMA)
Turki, Nagham Shakir& Mazal, Ilham Naghimish. Assessment of long distance chasing photometer (NAG-ADF-300-2) by estimating the drug atenolol with ammonium molybdate via continuous flow injection analysis. Baghdad Science Journal. 2020. Vol. 17, no. 1, pp.78-92.
https://search.emarefa.net/detail/BIM-935378
Data Type
Journal Articles
Language
English
Notes
Includes bibliographical references : p. 90-96
Record ID
BIM-935378