Detection Efficiency of NaI(Tl)‎ Detector in 511–1332 keV Energy Range

Joint Authors

Akkurt, I.
Gunoglu, K.
Arda, S. S.

Source

Science and Technology of Nuclear Installations

Issue

Vol. 2014, Issue 2014 (31 Dec. 2014), pp.1-5, 5 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2014-03-12

Country of Publication

Egypt

No. of Pages

5

Abstract EN

As it is important to obtain accurate analytical result in an experimental research, this required quality control of the experimental system.

Gamma spectrometry system can be used in a variety of different fields such as radiation and medical physics.

In this paper the absolute efficiency, peak to valley ratio, and energy resolution of a 3 ′ ′ × 3 ′ ′ NaI(Tl) detector were determined experimentally for 511, 662, 835, 1173, 1275, and 1332 keV photon energies obtained from 22Na, 54Mn, 60Co, and 137Cs radioactive sources.

American Psychological Association (APA)

Akkurt, I.& Gunoglu, K.& Arda, S. S.. 2014. Detection Efficiency of NaI(Tl) Detector in 511–1332 keV Energy Range. Science and Technology of Nuclear Installations،Vol. 2014, no. 2014, pp.1-5.
https://search.emarefa.net/detail/BIM-1047730

Modern Language Association (MLA)

Akkurt, I.…[et al.]. Detection Efficiency of NaI(Tl) Detector in 511–1332 keV Energy Range. Science and Technology of Nuclear Installations No. 2014 (2014), pp.1-5.
https://search.emarefa.net/detail/BIM-1047730

American Medical Association (AMA)

Akkurt, I.& Gunoglu, K.& Arda, S. S.. Detection Efficiency of NaI(Tl) Detector in 511–1332 keV Energy Range. Science and Technology of Nuclear Installations. 2014. Vol. 2014, no. 2014, pp.1-5.
https://search.emarefa.net/detail/BIM-1047730

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID