The AMBU® Aura-i™ Laryngeal Mask and LMA Supreme™: A Randomized Trial of Clinical Performance and Fibreoptic Positioning in Unparalysed, Anaesthetised Patients by Novices

Joint Authors

Yahaya, Zanahriah
Teoh, Wendy H.
Dintan, Nora A.
Agrawal, Ravi

Source

Anesthesiology Research and Practice

Issue

Vol. 2016, Issue 2016 (31 Dec. 2016), pp.1-8, 8 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2016-10-25

Country of Publication

Egypt

No. of Pages

8

Main Subjects

Medicine

Abstract EN

Introduction.

Manikin studies’ data cannot accurately be extrapolated to real-life scenarios and inherent differences in design and materials of newer products may affect their clinical performance.

Methods.

Hence, we compared the AMBU® Aura-i™ and LMA Supreme™ in this randomized trial involving 100 ASA 1-2 unparalysed anaesthetised patients undergoing minor gynaecological surgery.

Investigators had <20 Aura-i insertions.

Primary outcome was time to achieve effective ventilation and secondarily insertion parameters, oropharyngeal leak pressures (OLP), fibreoptic positioning, and pharyngeal morbidity.

The position of the Ambu Aura-i was evaluated with the Ascope; the fiberoptic view of the glottis was scored on a five-point scale.

Results.

43 (86%) AMBU Aura-i and 44 (88%) LMA Supremes were successfully inserted on first attempt (p=0.59), with similar ease (p=0.79), and comparable times to first capnogram, mean (SD) 18.2 (6.0) versus 17.3 (6.4) sec, p=0.9.

The Aura-i needed significantly less volume of air to inflate its cuff to 60 cmH2O on the manometer, 17.7 (3.5) versus 23.1 (4.4) mL, p<0.001.

Both devices exhibited similar OLP, Aura-i versus LMA Supreme, mean (SD) 28.8 (7.1) versus 27.3 (5.3) cmH2O, p=0.24.

There was no difference in ease of insertion or adjustment manoeuvres to aid ventilation.

90% of patients had good positioning of Aura-i on fibreoptic check, yielding a view of the vocal cords and epiglottis.

In 5 patients (10%), the vocal cords were not seen, but ventilatory function was unaffected.

Conclusions.

The Aura-i handled well in novices hands, with comparable times to insert and establish ventilation, similar leak pressures, and successful first attempt insertion rates compared to the LMA Supreme.

American Psychological Association (APA)

Yahaya, Zanahriah& Teoh, Wendy H.& Dintan, Nora A.& Agrawal, Ravi. 2016. The AMBU® Aura-i™ Laryngeal Mask and LMA Supreme™: A Randomized Trial of Clinical Performance and Fibreoptic Positioning in Unparalysed, Anaesthetised Patients by Novices. Anesthesiology Research and Practice،Vol. 2016, no. 2016, pp.1-8.
https://search.emarefa.net/detail/BIM-1096669

Modern Language Association (MLA)

Yahaya, Zanahriah…[et al.]. The AMBU® Aura-i™ Laryngeal Mask and LMA Supreme™: A Randomized Trial of Clinical Performance and Fibreoptic Positioning in Unparalysed, Anaesthetised Patients by Novices. Anesthesiology Research and Practice No. 2016 (2016), pp.1-8.
https://search.emarefa.net/detail/BIM-1096669

American Medical Association (AMA)

Yahaya, Zanahriah& Teoh, Wendy H.& Dintan, Nora A.& Agrawal, Ravi. The AMBU® Aura-i™ Laryngeal Mask and LMA Supreme™: A Randomized Trial of Clinical Performance and Fibreoptic Positioning in Unparalysed, Anaesthetised Patients by Novices. Anesthesiology Research and Practice. 2016. Vol. 2016, no. 2016, pp.1-8.
https://search.emarefa.net/detail/BIM-1096669

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-1096669