The Effects of Sample Transport by Pneumatic Tube System on Routine Hematology and Coagulation Tests

Joint Authors

Subbarayan, Devi
Choccalingam, Chidambharam
Lakshmi, Chittode Kodumudi Anantha

Source

Advances in Hematology

Issue

Vol. 2018, Issue 2018 (31 Dec. 2018), pp.1-4, 4 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2018-07-02

Country of Publication

Egypt

No. of Pages

4

Main Subjects

Diseases
Medicine

Abstract EN

Background.

Automation helps improve laboratory operational efficiency and reduce the turnaround time.

Pneumatic tube systems (PTS) automate specimen transport between the lab and other areas of the hospital.

Its effect on complete blood count (CBC) and coagulation is still controversial.

Aim.

To study the effects of pneumatic tube system sample transport on complete blood count and coagulation parameters to compare them with hand delivered samples.

Methods.

75 paired samples for complete blood count and 25 paired samples for coagulation analysis were compared between samples sent via pneumatic tube system and hand delivered system.

Results.

PTS showed significant decrease in red cell indices such as MCV and RDW and increase in MCHC.

Other red cell parameters and WBC parameters showed no statistical significant difference.

Statistically significant increase in platelet count was observed with PTS samples.

However, these differences were clinically insignificant.

No significant effect of PTS was found in PT and APTT samples compared to the hand delivered samples.

Conclusion.

Despite statistically significant changes in RBC parameters such as MCV, RDW, and MCHC and platelet count, these changes were clinically insignificant.

Hence, blood samples for CBC and coagulation assay can safely be transported via our hospital’s PTS.

However, further studies on platelet count are warranted to ensure safe transport and accuracy of the results.

American Psychological Association (APA)

Subbarayan, Devi& Choccalingam, Chidambharam& Lakshmi, Chittode Kodumudi Anantha. 2018. The Effects of Sample Transport by Pneumatic Tube System on Routine Hematology and Coagulation Tests. Advances in Hematology،Vol. 2018, no. 2018, pp.1-4.
https://search.emarefa.net/detail/BIM-1117753

Modern Language Association (MLA)

Subbarayan, Devi…[et al.]. The Effects of Sample Transport by Pneumatic Tube System on Routine Hematology and Coagulation Tests. Advances in Hematology No. 2018 (2018), pp.1-4.
https://search.emarefa.net/detail/BIM-1117753

American Medical Association (AMA)

Subbarayan, Devi& Choccalingam, Chidambharam& Lakshmi, Chittode Kodumudi Anantha. The Effects of Sample Transport by Pneumatic Tube System on Routine Hematology and Coagulation Tests. Advances in Hematology. 2018. Vol. 2018, no. 2018, pp.1-4.
https://search.emarefa.net/detail/BIM-1117753

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-1117753