Use of Tekscan K-Scan Sensors for Retropatellar Pressure Measurement Avoiding Errors during Implantation and the Effects of Shear Forces on the Measurement Precision

المؤلفون المشاركون

Wilharm, A.
Hurschler, Ch.
Dermitas, T.
Bohnsack, M.

المصدر

BioMed Research International

العدد

المجلد 2013، العدد 2013 (31 ديسمبر/كانون الأول 2013)، ص ص. 1-7، 7ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2013-12-03

دولة النشر

مصر

عدد الصفحات

7

التخصصات الرئيسية

الطب البشري

الملخص EN

Pressure-sensitive K-Scan 4000 sensors (Tekscan, USA) provide new possibilities for the dynamic measurement of force and pressure in biomechanical investigations.

We examined the sensors to determine in particular whether they are also suitable for reliable measurements of retropatellar forces and pressures.

Insertion approaches were also investigated and a lateral parapatellar arthrotomy supplemented by parapatellar sutures proved to be the most reliable method.

The ten human cadaver knees were tested in a knee-simulating machine at a torque of 30 and 40 Nm.

Each test cycle involved a dynamic extension from 120° flexion.

All recorded parameters showed a decrease of 1-2% per measurement cycle.

Although we supplemented the sensors with a Teflon film, the decrease, which was likely caused by shear force, was significant.

We evaluated 12 cycles and observed a linear decrease in parameters up to 17.2% (coefficient of regression 0.69–0.99).

In our opinion, the linear decrease can be considered a systematic error and can therefore be quantified and accounted for in subsequent experiments.

That will ensure reliable retropatellar usage of Tekscan sensors and distinguish the effects of knee joint surgeries from sensor wear-related effects.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Wilharm, A.& Hurschler, Ch.& Dermitas, T.& Bohnsack, M.. 2013. Use of Tekscan K-Scan Sensors for Retropatellar Pressure Measurement Avoiding Errors during Implantation and the Effects of Shear Forces on the Measurement Precision. BioMed Research International،Vol. 2013, no. 2013, pp.1-7.
https://search.emarefa.net/detail/BIM-1005214

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Wilharm, A.…[et al.]. Use of Tekscan K-Scan Sensors for Retropatellar Pressure Measurement Avoiding Errors during Implantation and the Effects of Shear Forces on the Measurement Precision. BioMed Research International No. 2013 (2013), pp.1-7.
https://search.emarefa.net/detail/BIM-1005214

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Wilharm, A.& Hurschler, Ch.& Dermitas, T.& Bohnsack, M.. Use of Tekscan K-Scan Sensors for Retropatellar Pressure Measurement Avoiding Errors during Implantation and the Effects of Shear Forces on the Measurement Precision. BioMed Research International. 2013. Vol. 2013, no. 2013, pp.1-7.
https://search.emarefa.net/detail/BIM-1005214

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1005214