Comparison of International Normalized Ratio Measurement between CoaguChek XS Plus and STA-R Coagulation Analyzers
المؤلفون المشاركون
La Gioia, Antonio
Kim, Hanah
Park, Chul Min
Choi, Sang-Gyu
Choi, Ju-Hee
Moon, Hee-Won
Yun, Yeo-Min
Hur, Mina
المصدر
العدد
المجلد 2013، العدد 2013 (31 ديسمبر/كانون الأول 2013)، ص ص. 1-6، 6ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2012-12-24
دولة النشر
مصر
عدد الصفحات
6
التخصصات الرئيسية
الملخص EN
Background.
Point-of-care testing (POCT) coagulometers are increasingly being used in the hospital setting.
We investigated whether the prothrombin time international normalized ratio (INR) results by CoaguChek XS Plus (Roche Diagnostics GmbH, Mannheim, Germany) can be used reliably without being confirmed with the INR results by STA-R system (Diagnostica Stago S.A.S, Asnières sur Seine, France).
Methods.
A total of 118 INR measurements by CoaguChek XS Plus and STA-R were compared using Passing/Bablok regression analysis and Bland-Altman plot.
Agreement of the INR measurements was further assessed in relation to dosing decision.
Results.
The correlation of INR measurements between CoaguChek XS Plus and STA-R was excellent (correlation coefficient = 0.964).
The mean difference tended to increase as INR results increased and was 0.25 INR in the therapeutic range (2.0-3.0 INR).
The overall agreement was fair to good (kappa = 0.679), and 21/118 (17.8%) INR measurements showed a difference in dosing decision.
Conclusion.
The positive bias of CoaguChek XS Plus may be obvious even in the therapeutic INR range, and dosing decision based on the CoaguChek XS Plus INR results would be different from that based on the STA-R results.
The INR measurements by POCT coagulometers still need to be confirmed with the laboratory INR measurements.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Hur, Mina& Kim, Hanah& Park, Chul Min& La Gioia, Antonio& Choi, Sang-Gyu& Choi, Ju-Hee…[et al.]. 2012. Comparison of International Normalized Ratio Measurement between CoaguChek XS Plus and STA-R Coagulation Analyzers. BioMed Research International،Vol. 2013, no. 2013, pp.1-6.
https://search.emarefa.net/detail/BIM-1030205
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Hur, Mina…[et al.]. Comparison of International Normalized Ratio Measurement between CoaguChek XS Plus and STA-R Coagulation Analyzers. BioMed Research International No. 2013 (2013), pp.1-6.
https://search.emarefa.net/detail/BIM-1030205
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Hur, Mina& Kim, Hanah& Park, Chul Min& La Gioia, Antonio& Choi, Sang-Gyu& Choi, Ju-Hee…[et al.]. Comparison of International Normalized Ratio Measurement between CoaguChek XS Plus and STA-R Coagulation Analyzers. BioMed Research International. 2012. Vol. 2013, no. 2013, pp.1-6.
https://search.emarefa.net/detail/BIM-1030205
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-1030205
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر