Comparison of International Normalized Ratio Measurement between CoaguChek XS Plus and STA-R Coagulation Analyzers

المؤلفون المشاركون

La Gioia, Antonio
Kim, Hanah
Park, Chul Min
Choi, Sang-Gyu
Choi, Ju-Hee
Moon, Hee-Won
Yun, Yeo-Min
Hur, Mina

المصدر

BioMed Research International

العدد

المجلد 2013، العدد 2013 (31 ديسمبر/كانون الأول 2013)، ص ص. 1-6، 6ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2012-12-24

دولة النشر

مصر

عدد الصفحات

6

التخصصات الرئيسية

الطب البشري

الملخص EN

Background.

Point-of-care testing (POCT) coagulometers are increasingly being used in the hospital setting.

We investigated whether the prothrombin time international normalized ratio (INR) results by CoaguChek XS Plus (Roche Diagnostics GmbH, Mannheim, Germany) can be used reliably without being confirmed with the INR results by STA-R system (Diagnostica Stago S.A.S, Asnières sur Seine, France).

Methods.

A total of 118 INR measurements by CoaguChek XS Plus and STA-R were compared using Passing/Bablok regression analysis and Bland-Altman plot.

Agreement of the INR measurements was further assessed in relation to dosing decision.

Results.

The correlation of INR measurements between CoaguChek XS Plus and STA-R was excellent (correlation coefficient = 0.964).

The mean difference tended to increase as INR results increased and was 0.25 INR in the therapeutic range (2.0-3.0 INR).

The overall agreement was fair to good (kappa = 0.679), and 21/118 (17.8%) INR measurements showed a difference in dosing decision.

Conclusion.

The positive bias of CoaguChek XS Plus may be obvious even in the therapeutic INR range, and dosing decision based on the CoaguChek XS Plus INR results would be different from that based on the STA-R results.

The INR measurements by POCT coagulometers still need to be confirmed with the laboratory INR measurements.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Hur, Mina& Kim, Hanah& Park, Chul Min& La Gioia, Antonio& Choi, Sang-Gyu& Choi, Ju-Hee…[et al.]. 2012. Comparison of International Normalized Ratio Measurement between CoaguChek XS Plus and STA-R Coagulation Analyzers. BioMed Research International،Vol. 2013, no. 2013, pp.1-6.
https://search.emarefa.net/detail/BIM-1030205

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Hur, Mina…[et al.]. Comparison of International Normalized Ratio Measurement between CoaguChek XS Plus and STA-R Coagulation Analyzers. BioMed Research International No. 2013 (2013), pp.1-6.
https://search.emarefa.net/detail/BIM-1030205

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Hur, Mina& Kim, Hanah& Park, Chul Min& La Gioia, Antonio& Choi, Sang-Gyu& Choi, Ju-Hee…[et al.]. Comparison of International Normalized Ratio Measurement between CoaguChek XS Plus and STA-R Coagulation Analyzers. BioMed Research International. 2012. Vol. 2013, no. 2013, pp.1-6.
https://search.emarefa.net/detail/BIM-1030205

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1030205