Estimation of Measurement Uncertainties for the DGT Passive Sampler Used for Determination of Copper in Water

المؤلفون المشاركون

Knutsson, Jesper
Rauch, Sebastien
Morrison, Gregory M.

المصدر

International Journal of Analytical Chemistry

العدد

المجلد 2014، العدد 2014 (31 ديسمبر/كانون الأول 2014)، ص ص. 1-7، 7ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2014-09-01

دولة النشر

مصر

عدد الصفحات

7

التخصصات الرئيسية

الكيمياء
علوم

الملخص EN

Diffusion-based passive samplers are increasingly used for water quality monitoring.

While the overall method robustness and reproducibility for passive samplers in water are widely reported, there has been a lack of a detailed description of uncertainty sources.

In this paper an uncertainty budget for the determination of fully labile Cu in water using a DGT passive sampler is presented.

Uncertainty from the estimation of effective cross-sectional diffusion area and the instrumental determination of accumulated mass of analyte are the most significant sources of uncertainty, while uncertainties from contamination and the estimation of diffusion coefficient are negligible.

The results presented highlight issues with passive samplers which are important to address if overall method uncertainty is to be reduced and effective strategies to reduce overall method uncertainty are presented.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Knutsson, Jesper& Rauch, Sebastien& Morrison, Gregory M.. 2014. Estimation of Measurement Uncertainties for the DGT Passive Sampler Used for Determination of Copper in Water. International Journal of Analytical Chemistry،Vol. 2014, no. 2014, pp.1-7.
https://search.emarefa.net/detail/BIM-1035932

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Knutsson, Jesper…[et al.]. Estimation of Measurement Uncertainties for the DGT Passive Sampler Used for Determination of Copper in Water. International Journal of Analytical Chemistry No. 2014 (2014), pp.1-7.
https://search.emarefa.net/detail/BIM-1035932

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Knutsson, Jesper& Rauch, Sebastien& Morrison, Gregory M.. Estimation of Measurement Uncertainties for the DGT Passive Sampler Used for Determination of Copper in Water. International Journal of Analytical Chemistry. 2014. Vol. 2014, no. 2014, pp.1-7.
https://search.emarefa.net/detail/BIM-1035932

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1035932