Comparison of Anterior Corneal Curvature Measurements Using a Galilei Dual Scheimpflug Analyzer and Topcon Auto Kerato-Refractometer
المؤلفون المشاركون
Wang, Xiaogang
Dong, Jing
Wu, Qiang
المصدر
العدد
المجلد 2014، العدد 2014 (31 ديسمبر/كانون الأول 2014)، ص ص. 1-5، 5ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2014-05-22
دولة النشر
مصر
عدد الصفحات
5
التخصصات الرئيسية
الملخص EN
Purpose.
To compare anterior corneal keratometry (K) measurements taken by a dual Scheimpflug analyzer and an auto kerato-refractometer.
Methods.
Sixty-four normal eyes underwent keratometric measurements with both devices.
The repeatability of the auto kerato-refractometer measurements was assessed by calculating the coefficient of variation (COV).
The interdevice agreement was evaluated using the Bland-Altman analysis, Pearson correlation coefficient, and paired two-tailed t -test.
Results.
The COV of the flat K and steep K measurements taken by the auto kerato-refractometer were 0.21% and 0.29%, respectively.
There were no significant differences between the steep K and average K measurements for the Topcon and Galilei devices ( P = 0.475, and P = 0.137, resp.).
The Galilei flat K values were lower than those of the Topcon ( P = 0.002).
Both of the instruments showed good agreement for all anterior corneal keratometric values.
There was a significant linear correlation between the Galilei and Topcon devices for the flat K ( r = 0.989, P < 0.0001), steep K ( r = 0.987, P < 0.0001), and average K values ( r = 0.994, P < 0.0001).
Conclusions.
The anterior corneal flat keratometric values were not interchangeable between the Galilei and Topcon devices.
However, the measurements of the two instruments showed significant linear correlation with each other.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Wang, Xiaogang& Dong, Jing& Wu, Qiang. 2014. Comparison of Anterior Corneal Curvature Measurements Using a Galilei Dual Scheimpflug Analyzer and Topcon Auto Kerato-Refractometer. Journal of Ophthalmology،Vol. 2014, no. 2014, pp.1-5.
https://search.emarefa.net/detail/BIM-1042408
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Wang, Xiaogang…[et al.]. Comparison of Anterior Corneal Curvature Measurements Using a Galilei Dual Scheimpflug Analyzer and Topcon Auto Kerato-Refractometer. Journal of Ophthalmology No. 2014 (2014), pp.1-5.
https://search.emarefa.net/detail/BIM-1042408
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Wang, Xiaogang& Dong, Jing& Wu, Qiang. Comparison of Anterior Corneal Curvature Measurements Using a Galilei Dual Scheimpflug Analyzer and Topcon Auto Kerato-Refractometer. Journal of Ophthalmology. 2014. Vol. 2014, no. 2014, pp.1-5.
https://search.emarefa.net/detail/BIM-1042408
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-1042408
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر