Corneal Biomechanical Assessment Using Corneal Visualization Scheimpflug Technology in Keratoconic and Normal Eyes
المؤلفون المشاركون
Tian, Lei
Huang, Yi-Fei
Wang, Li-Qiang
Bai, Hua
Wang, Qun
Jiang, Jing-Jing
Wu, Ying
Gao, Min
المصدر
العدد
المجلد 2014، العدد 2014 (31 ديسمبر/كانون الأول 2014)، ص ص. 1-8، 8ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2014-03-30
دولة النشر
مصر
عدد الصفحات
8
التخصصات الرئيسية
الملخص EN
Purpose.
To compare the corneal biomechanical properties of keratoconic patients and age-matched controls using corneal visualization Scheimpflug technology (Corvis ST).
Methods.
Sixty keratoconic eyes from 47 keratoconus patients and 60 normal eyes from 60 controls were enrolled in this prospective study.
Tomography and biomechanical parameters of all eyes were obtained with the Pentacam and Corvis ST, respectively.
Intraocular pressure was measured using a Goldmann applanation tonometer.
Results.
The tomography and biomechanical parameters of the keratoconic corneas were significantly different from those of the normal corneas except for the anterior chamber angle, first applanation length, the highest concavity time, and peak distance.
The deformation amplitude was the best predictive parameter (area under the curve: 0.882), with a sensitivity of 81.7%, although there was a significant overlap between keratoconic and normal corneas that ranged from 1.0 to 1.4 mm.
In both the keratoconus and control groups, the deformation amplitude was negatively correlated with intraocular pressure, central corneal thickness, and corneal volume at 3 and 5 mm.
Conclusions.
Corvis ST offers an alternative method for measuring corneal biomechanical properties.
The possibility of classifying keratoconus based on deformation amplitude deserves clinical attention.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Tian, Lei& Huang, Yi-Fei& Wang, Li-Qiang& Bai, Hua& Wang, Qun& Jiang, Jing-Jing…[et al.]. 2014. Corneal Biomechanical Assessment Using Corneal Visualization Scheimpflug Technology in Keratoconic and Normal Eyes. Journal of Ophthalmology،Vol. 2014, no. 2014, pp.1-8.
https://search.emarefa.net/detail/BIM-1042410
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Tian, Lei…[et al.]. Corneal Biomechanical Assessment Using Corneal Visualization Scheimpflug Technology in Keratoconic and Normal Eyes. Journal of Ophthalmology No. 2014 (2014), pp.1-8.
https://search.emarefa.net/detail/BIM-1042410
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Tian, Lei& Huang, Yi-Fei& Wang, Li-Qiang& Bai, Hua& Wang, Qun& Jiang, Jing-Jing…[et al.]. Corneal Biomechanical Assessment Using Corneal Visualization Scheimpflug Technology in Keratoconic and Normal Eyes. Journal of Ophthalmology. 2014. Vol. 2014, no. 2014, pp.1-8.
https://search.emarefa.net/detail/BIM-1042410
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-1042410
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
![](/images/ebook-kashef.png)
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر
![](/images/kashef-image.png)