Purpose.

To assess the repeatability and reproducibility of ocular biometry and intraocular lens (IOL) power measurements obtained by ophthalmology residents using an AL-Scan device, a novel optical biometer.

Methods.

Two ophthalmology residents were instructed regarding the AL-Scan device.

Both performed ocular biometry and IOL power measurements using AL-Scan, three times on each of 128 eyes, independently of one another.

Corneal keratometry readings, horizontal iris width, central corneal thickness, anterior chamber depth, pupil size, and axial length values measured by both residents were recorded together with IOL power values calculated on the basis of four different IOL calculation formulas (SRK/T, Holladay, and HofferQ).

Repeatability and reproducibility of the measurements obtained were analyzed using the intraclass correlation coefficient (ICC).

Results.

Repeatability (ICC, 0.872-0.999 for resident 1 versus 0.905-0.999 for resident 2) and reproducibility (ICC, 0.916-0.999) were high for all biometric measurements.

Repeatability (ICC, 0.981-0.983 for resident 1 versus 0.995-0.996 for resident 2) and reproducibility were also high for all IOL power measurements (ICC, 0.996 for all).

Conclusions.

The AL-Scan device exhibits good repeatability and reproducibility in all biometric measurements and IOL power calculations, independent of the operator concerned.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Kola, Mehmet& Duran, Hikmet& Turk, Adem& Mollamehmetoglu, Suleyman& Kalkisim, Ahmet& Erdol, Hidayet. 2014. Evaluation of the Repeatability and the Reproducibility of AL-Scan Measurements Obtained by Residents. Journal of Ophthalmology،Vol. 2014, no. 2014, pp.1-6.
https://search.emarefa.net/detail/BIM-1042608

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Kola, Mehmet…[et al.]. Evaluation of the Repeatability and the Reproducibility of AL-Scan Measurements Obtained by Residents. Journal of Ophthalmology No. 2014 (2014), pp.1-6.
https://search.emarefa.net/detail/BIM-1042608

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Kola, Mehmet& Duran, Hikmet& Turk, Adem& Mollamehmetoglu, Suleyman& Kalkisim, Ahmet& Erdol, Hidayet. Evaluation of the Repeatability and the Reproducibility of AL-Scan Measurements Obtained by Residents. Journal of Ophthalmology. 2014. Vol. 2014, no. 2014, pp.1-6.
https://search.emarefa.net/detail/BIM-1042608

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1042608