Design and Evaluation of a Fully Implantable Control Unit for Blood Pumps

المؤلفون المشاركون

Unthan, Kristin
Gräf, Felix
Laumen, Marco
Finocchiaro, Thomas
Sommer, Christoph
Lanmüller, Hermann
Steinseifer, Ulrich

المصدر

BioMed Research International

العدد

المجلد 2015، العدد 2015 (31 ديسمبر/كانون الأول 2015)، ص ص. 1-9، 9ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2015-10-25

دولة النشر

مصر

عدد الصفحات

9

التخصصات الرئيسية

الطب البشري

الملخص EN

As the number of donor hearts is limited while more and more patients suffer from end stage biventricular heart failure, Total Artificial Hearts become a promising alternative to conventional treatment.

While pneumatic devices sufficiently supply the patients with blood flow, the patient’s quality of life is limited by the percutaneous pressure lines and the size of the external control unit.

This paper describes the development of the control unit of the ReinHeart, a fully implantable Total Artificial Heart.

General requirements for any implantable control unit are defined from a technical and medical point of view: necessity of a Transcutaneous Energy Transmission, autonomous operation, safety, geometry, and efficiency.

Based on the requirements, a prototype is designed; it incorporates a LiFePo4 battery pack with charger, a rectifier for transcutaneous energy transmission, the motor’s driver electronics, and a microcontroller which monitors and controls all functions.

In validation tests, the control unit demonstrated a stable operation on TET and battery supply and a safe switching from one supply to the other.

The overall mean efficiency is 14% on TET and 22% on battery supply.

The control unit is suitable for chronic animal trials of the ReinHeart.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Unthan, Kristin& Gräf, Felix& Laumen, Marco& Finocchiaro, Thomas& Sommer, Christoph& Lanmüller, Hermann…[et al.]. 2015. Design and Evaluation of a Fully Implantable Control Unit for Blood Pumps. BioMed Research International،Vol. 2015, no. 2015, pp.1-9.
https://search.emarefa.net/detail/BIM-1054780

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Unthan, Kristin…[et al.]. Design and Evaluation of a Fully Implantable Control Unit for Blood Pumps. BioMed Research International No. 2015 (2015), pp.1-9.
https://search.emarefa.net/detail/BIM-1054780

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Unthan, Kristin& Gräf, Felix& Laumen, Marco& Finocchiaro, Thomas& Sommer, Christoph& Lanmüller, Hermann…[et al.]. Design and Evaluation of a Fully Implantable Control Unit for Blood Pumps. BioMed Research International. 2015. Vol. 2015, no. 2015, pp.1-9.
https://search.emarefa.net/detail/BIM-1054780

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1054780