Assessment of Random Error in Phantom Dosimetry with the Use of Error Simulation in Statistical Software
المؤلفون المشاركون
Hoogeveen, R. C.
Martens, E. P.
van der Stelt, P. F.
Berkhout, W. E. R.
المصدر
العدد
المجلد 2015، العدد 2015 (31 ديسمبر/كانون الأول 2015)، ص ص. 1-5، 5ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2015-12-31
دولة النشر
مصر
عدد الصفحات
5
التخصصات الرئيسية
الملخص EN
Objective.
To investigate if software simulation is practical for quantifying random error (RE) in phantom dosimetry.
Materials and Methods.
We applied software error simulation to an existing dosimetry study.
The specifications and the measurement values of this study were brought into the software (R version 3.0.2) together with the algorithm of the calculation of the effective dose (E).
Four sources of RE were specified: (1) the calibration factor; (2) the background radiation correction; (3) the read-out process of the dosimeters; and (4) the fluctuation of the X-ray generator.
Results.
The amount of RE introduced by these sources was calculated on the basis of the experimental values and the mathematical rules of error propagation.
The software repeated the calculations of E multiple times (n=10,000) while attributing the applicable RE to the experimental values.
A distribution of E emerged as a confidence interval around an expected value.
Conclusions.
Credible confidence intervals around E in phantom dose studies can be calculated by using software modelling of the experiment.
With credible confidence intervals, the statistical significance of differences between protocols can be substantiated or rejected.
This modelling software can also be used for a power analysis when planning phantom dose experiments.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Hoogeveen, R. C.& Martens, E. P.& van der Stelt, P. F.& Berkhout, W. E. R.. 2015. Assessment of Random Error in Phantom Dosimetry with the Use of Error Simulation in Statistical Software. BioMed Research International،Vol. 2015, no. 2015, pp.1-5.
https://search.emarefa.net/detail/BIM-1056066
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Hoogeveen, R. C.…[et al.]. Assessment of Random Error in Phantom Dosimetry with the Use of Error Simulation in Statistical Software. BioMed Research International No. 2015 (2015), pp.1-5.
https://search.emarefa.net/detail/BIM-1056066
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Hoogeveen, R. C.& Martens, E. P.& van der Stelt, P. F.& Berkhout, W. E. R.. Assessment of Random Error in Phantom Dosimetry with the Use of Error Simulation in Statistical Software. BioMed Research International. 2015. Vol. 2015, no. 2015, pp.1-5.
https://search.emarefa.net/detail/BIM-1056066
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-1056066
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر