Rapid Switch from Intra-Aortic Balloon Pumping to Percutaneous Cardiopulmonary Support Using Perclose ProGlide

المؤلفون المشاركون

Momomura, Shin-ichi
Sakakura, Kenichi
Adachi, Yusuke
Taniguchi, Yousuke
Wada, Hiroshi
Fujita, Hideo

المصدر

Case Reports in Cardiology

العدد

المجلد 2015، العدد 2015 (31 ديسمبر/كانون الأول 2015)، ص ص. 1-5، 5ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2015-12-09

دولة النشر

مصر

عدد الصفحات

5

التخصصات الرئيسية

الأمراض

الملخص EN

We present a case of a patient who needed rapid switch from intra-aortic balloon pumping (IABP) to percutaneous cardiopulmonary support (PCPS)/venoarterial extracorporeal membrane oxygenation.

It is difficult to switch from IABP to PCPS, because 0.035-inch guidewires cannot pass the IABP guidewire lumen (0.025-inch compatible), and the IABP sheath needs to be removed together with the IABP catheter.

First, a 0.025-inch guidewire was inserted into the IABP wire lumen, and then the IABP catheter together with the 8 Fr IABP sheath was removed, leaving the 0.025-inch guidewire in place.

We used the Perclose ProGlide for safe and rapid exchange of the 0.025-inch guidewire for a 0.035-inch guidewire.

This allowed insertion of a PCPS cannula and the prompt initiation of PCPS.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Sakakura, Kenichi& Adachi, Yusuke& Taniguchi, Yousuke& Wada, Hiroshi& Momomura, Shin-ichi& Fujita, Hideo. 2015. Rapid Switch from Intra-Aortic Balloon Pumping to Percutaneous Cardiopulmonary Support Using Perclose ProGlide. Case Reports in Cardiology،Vol. 2015, no. 2015, pp.1-5.
https://search.emarefa.net/detail/BIM-1058112

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Sakakura, Kenichi…[et al.]. Rapid Switch from Intra-Aortic Balloon Pumping to Percutaneous Cardiopulmonary Support Using Perclose ProGlide. Case Reports in Cardiology No. 2015 (2015), pp.1-5.
https://search.emarefa.net/detail/BIM-1058112

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Sakakura, Kenichi& Adachi, Yusuke& Taniguchi, Yousuke& Wada, Hiroshi& Momomura, Shin-ichi& Fujita, Hideo. Rapid Switch from Intra-Aortic Balloon Pumping to Percutaneous Cardiopulmonary Support Using Perclose ProGlide. Case Reports in Cardiology. 2015. Vol. 2015, no. 2015, pp.1-5.
https://search.emarefa.net/detail/BIM-1058112

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1058112