Comparison of Measurement of Central Corneal Thickness with Spectral Domain Optical Coherence Tomography and Standard Ultrasonic Pachymeter in Premature Infants

المؤلفون المشاركون

Erol, Muhammet Kazim
Turgut Coban, Deniz
Dogan, Berna
Toslak, Devrim
Hekimoglu, Emre
Yucel, Ozgur

المصدر

Journal of Ophthalmology

العدد

المجلد 2015، العدد 2015 (31 ديسمبر/كانون الأول 2015)، ص ص. 1-6، 6ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2015-07-22

دولة النشر

مصر

عدد الصفحات

6

التخصصات الرئيسية

الطب البشري

الملخص EN

Purpose.

To evaluate the repeatability of measurement of central corneal thickness (CCT) by spectral domain optical coherence (SD-OCT) in premature infants and compare it to CCT measurement by ultrasonic pachymetry (USP).

Methods.

Three CCT measurements of the left eyes of 50 premature infants were obtained by SD-OCT using the iVue system.

10 CCT measurements of each 28 left eyes of 28 infants were obtained by USP using the Pacscan 300P system.

Bland-Altman plots were developed and the limit of agreement (LoA) was determined to compare the mean of the SD-OCT and USP measurements.

Results.

No statistically significant difference was found among the 3 CCT measurements by SD-OCT.

Both USP and SD-OCT have been performed for only left eyes of 28 of the 50 babies.

Those results have been compared with each other.

A statistically significant difference was found between the mean CCT measurements by SD-OCT and USP (p<0.05).

The LoA between the SD-OCT and USP measurements ranged from 11.4 to −64.1.

Conclusions.

CCT can be measured using the iVue SD-OCT system with a high level of repeatability.

Although measurement of CCT by SD-OCT and USP is highly correlated, the 2 systems cannot be used interchangeably in premature infants.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Hekimoglu, Emre& Erol, Muhammet Kazim& Toslak, Devrim& Turgut Coban, Deniz& Dogan, Berna& Yucel, Ozgur. 2015. Comparison of Measurement of Central Corneal Thickness with Spectral Domain Optical Coherence Tomography and Standard Ultrasonic Pachymeter in Premature Infants. Journal of Ophthalmology،Vol. 2015, no. 2015, pp.1-6.
https://search.emarefa.net/detail/BIM-1069641

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Hekimoglu, Emre…[et al.]. Comparison of Measurement of Central Corneal Thickness with Spectral Domain Optical Coherence Tomography and Standard Ultrasonic Pachymeter in Premature Infants. Journal of Ophthalmology No. 2015 (2015), pp.1-6.
https://search.emarefa.net/detail/BIM-1069641

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Hekimoglu, Emre& Erol, Muhammet Kazim& Toslak, Devrim& Turgut Coban, Deniz& Dogan, Berna& Yucel, Ozgur. Comparison of Measurement of Central Corneal Thickness with Spectral Domain Optical Coherence Tomography and Standard Ultrasonic Pachymeter in Premature Infants. Journal of Ophthalmology. 2015. Vol. 2015, no. 2015, pp.1-6.
https://search.emarefa.net/detail/BIM-1069641

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1069641