Evaluation of Central Corneal Thickness Using Corneal Dynamic Scheimpflug Analyzer Corvis ST and Comparison with Pentacam Rotating Scheimpflug System and Ultrasound Pachymetry in Normal Eyes

المؤلفون المشاركون

Huang, Jinhai
Bao, Fangjun
Lu, Weicong
Wang, Qinmei
Yu, Ayong
Zhao, Weiqi
Huang, Zixu
Savini, G.

المصدر

Journal of Ophthalmology

العدد

المجلد 2015، العدد 2015 (31 ديسمبر/كانون الأول 2015)، ص ص. 1-8، 8ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2015-12-01

دولة النشر

مصر

عدد الصفحات

8

التخصصات الرئيسية

الطب البشري

الملخص EN

Purpose.

To assess the repeatability and reproducibility of central corneal thickness (CCT) measurements by corneal dynamic Scheimpflug analyzer Corvis ST in normal eyes and compare the agreement with Pentacam rotating Scheimpflug System and ultrasound pachymetry.

Methods.

84 right eyes underwent Corvis ST measurements performed by two operators.

The test-retest repeatability (TRT), within-subject coefficient of variation (CoV), and intraclass correlation coefficient (ICC) were used to evaluate the intraoperator repeatability and interoperator reproducibility.

CCT measurements also were obtained from Pentacam and ultrasound pachymetry by the first operator.

The agreement between the three devices was evaluated with 95% limits of agreement (LoA) and Bland-Altman plots.

Results.

Corvis ST showed high repeatability as indicated by TRT ≤ 13.0 μm, CoV < 0.9%, and ICC > 0.97.

The interoperator reproducibility was also excellent.

The CoV was <0.9%, and ICC was >0.97.

Corvis ST showed significantly lower values than Pentacam and ultrasound pachymetry ( P < 0.001 ).

The 95% LoA between Corvis ST and Pentacam or ultrasound pachymetry were −15.8 to 9.5 μm and −27.9 to 12.3 μm, respectively.

Conclusions.

Corvis ST showed excellent repeatability and interoperator reproducibility of CCT measurements in normal eyes.

Corvis ST is interchangeable with Pentacam but not with ultrasound pachymetry.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Yu, Ayong& Zhao, Weiqi& Savini, G.& Huang, Zixu& Bao, Fangjun& Lu, Weicong…[et al.]. 2015. Evaluation of Central Corneal Thickness Using Corneal Dynamic Scheimpflug Analyzer Corvis ST and Comparison with Pentacam Rotating Scheimpflug System and Ultrasound Pachymetry in Normal Eyes. Journal of Ophthalmology،Vol. 2015, no. 2015, pp.1-8.
https://search.emarefa.net/detail/BIM-1069834

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Yu, Ayong…[et al.]. Evaluation of Central Corneal Thickness Using Corneal Dynamic Scheimpflug Analyzer Corvis ST and Comparison with Pentacam Rotating Scheimpflug System and Ultrasound Pachymetry in Normal Eyes. Journal of Ophthalmology No. 2015 (2015), pp.1-8.
https://search.emarefa.net/detail/BIM-1069834

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Yu, Ayong& Zhao, Weiqi& Savini, G.& Huang, Zixu& Bao, Fangjun& Lu, Weicong…[et al.]. Evaluation of Central Corneal Thickness Using Corneal Dynamic Scheimpflug Analyzer Corvis ST and Comparison with Pentacam Rotating Scheimpflug System and Ultrasound Pachymetry in Normal Eyes. Journal of Ophthalmology. 2015. Vol. 2015, no. 2015, pp.1-8.
https://search.emarefa.net/detail/BIM-1069834

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1069834