![](/images/graphics-bg.png)
NaI(Tl) Detector Efficiency Computation Using Radioactive Parallelepiped Sources Based on Efficiency Transfer Principle
المؤلفون المشاركون
Badawi, Mohamed S.
Gouda, Mona M.
El-Khatib, Ahmed M.
Thabet, Abouzeid A.
Salim, Ahmed A.
Abbas, Mahmoud I.
المصدر
Science and Technology of Nuclear Installations
العدد
المجلد 2015، العدد 2015 (31 ديسمبر/كانون الأول 2015)، ص ص. 1-9، 9ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2015-11-12
دولة النشر
مصر
عدد الصفحات
9
الملخص EN
The efficiency transfer (ET) principle is considered as a simple numerical simulation method, which can be used to calculate the full-energy peak efficiency (FEPE) of 3″×3″ NaI(Tl) scintillation detector over a wide energy range.
In this work, the calculations of FEPE are based on computing the effective solid angle ratio between a radioactive point and parallelepiped sources located at various distances from the detector surface.
Besides, the attenuation of the photon by the source-to-detector system (detector material, detector end cap, and holder material) was considered and determined.
This method is straightforwardly useful in setting up the efficiency calibration curve for NaI(Tl) scintillation detector, when no calibration sources exist in volume shape.
The values of the efficiency calculations using theoretical method are compared with the measured ones and the results show that the discrepancies in general for all the measurements are found to be less than 6%.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Badawi, Mohamed S.& Gouda, Mona M.& El-Khatib, Ahmed M.& Thabet, Abouzeid A.& Salim, Ahmed A.& Abbas, Mahmoud I.. 2015. NaI(Tl) Detector Efficiency Computation Using Radioactive Parallelepiped Sources Based on Efficiency Transfer Principle. Science and Technology of Nuclear Installations،Vol. 2015, no. 2015, pp.1-9.
https://search.emarefa.net/detail/BIM-1077839
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Badawi, Mohamed S.…[et al.]. NaI(Tl) Detector Efficiency Computation Using Radioactive Parallelepiped Sources Based on Efficiency Transfer Principle. Science and Technology of Nuclear Installations No. 2015 (2015), pp.1-9.
https://search.emarefa.net/detail/BIM-1077839
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Badawi, Mohamed S.& Gouda, Mona M.& El-Khatib, Ahmed M.& Thabet, Abouzeid A.& Salim, Ahmed A.& Abbas, Mahmoud I.. NaI(Tl) Detector Efficiency Computation Using Radioactive Parallelepiped Sources Based on Efficiency Transfer Principle. Science and Technology of Nuclear Installations. 2015. Vol. 2015, no. 2015, pp.1-9.
https://search.emarefa.net/detail/BIM-1077839
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-1077839
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
![](/images/ebook-kashef.png)
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر
![](/images/kashef-image.png)