Differential Intracochlear Sound Pressure Measurements in Human Temporal Bones with an Off-the-Shelf Sensor

المؤلفون المشاركون

Lenarz, Thomas
Grossöhmichen, Martin
Salcher, Rolf
Püschel, Klaus
Maier, Hannes

المصدر

BioMed Research International

العدد

المجلد 2016، العدد 2016 (31 ديسمبر/كانون الأول 2016)، ص ص. 1-10، 10ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2016-08-16

دولة النشر

مصر

عدد الصفحات

10

التخصصات الرئيسية

الطب البشري

الملخص EN

The standard method to determine the output level of acoustic and mechanical stimulation to the inner ear is measurement of vibration response of the stapes in human cadaveric temporal bones (TBs) by laser Doppler vibrometry.

However, this method is reliable only if the intact ossicular chain is stimulated.

For other stimulation modes an alternative method is needed.

The differential intracochlear sound pressure between scala vestibuli (SV) and scala tympani (ST) is assumed to correlate with excitation.

Using a custom-made pressure sensor it has been successfully measured and used to determine the output level of acoustic and mechanical stimulation.

To make this method generally accessible, an off-the-shelf pressure sensor (Samba Preclin 420 LP, Samba Sensors) was tested here for intracochlear sound pressure measurements.

During acoustic stimulation, intracochlear sound pressures were simultaneously measurable in SV and ST between 0.1 and 8 kHz with sufficient signal-to-noise ratios with this sensor.

The pressure differences were comparable to results obtained with custom-made sensors.

Our results demonstrated that the pressure sensor Samba Preclin 420 LP is usable for measurements of intracochlear sound pressures in SV and ST and for the determination of differential intracochlear sound pressures.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Grossöhmichen, Martin& Salcher, Rolf& Püschel, Klaus& Lenarz, Thomas& Maier, Hannes. 2016. Differential Intracochlear Sound Pressure Measurements in Human Temporal Bones with an Off-the-Shelf Sensor. BioMed Research International،Vol. 2016, no. 2016, pp.1-10.
https://search.emarefa.net/detail/BIM-1098358

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Grossöhmichen, Martin…[et al.]. Differential Intracochlear Sound Pressure Measurements in Human Temporal Bones with an Off-the-Shelf Sensor. BioMed Research International No. 2016 (2016), pp.1-10.
https://search.emarefa.net/detail/BIM-1098358

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Grossöhmichen, Martin& Salcher, Rolf& Püschel, Klaus& Lenarz, Thomas& Maier, Hannes. Differential Intracochlear Sound Pressure Measurements in Human Temporal Bones with an Off-the-Shelf Sensor. BioMed Research International. 2016. Vol. 2016, no. 2016, pp.1-10.
https://search.emarefa.net/detail/BIM-1098358

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1098358