Volumetric Modulated Arc Therapy for Spine Radiosurgery: Superior Treatment Planning and Delivery Compared to Static Beam Intensity Modulated Radiotherapy

المؤلفون المشاركون

Harel, Ran
Zach, Leor
Tsvang, Lev
Alezra, Dror
Ben Ayun, Maoz

المصدر

BioMed Research International

العدد

المجلد 2016، العدد 2016 (31 ديسمبر/كانون الأول 2016)، ص ص. 1-6، 6ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2016-01-13

دولة النشر

مصر

عدد الصفحات

6

التخصصات الرئيسية

الطب البشري

الملخص EN

Purpose.

Spine stereotactic radiosurgery (SRS) delivers an accurate and efficient high radiation dose to vertebral metastases in 1–5 fractions.

We aimed to compare volumetric modulated arc therapy (VMAT) to static beam intensity modulated radiotherapy (IMRT) for spine SRS.

Methods and Materials.

Ten spine lesions of previously treated SRS patients were planned retrospectively using both IMRT and VMAT with a prescribed dose of 16 Gy to 100% of the planning target volume (PTV).

The plans were compared for conformity, homogeneity, treatment delivery time, and safety (spinal cord dose).

Results.

All evaluated parameters favored the VMAT plan over the IMRT plans.

D m i n in the IMRT was significantly lower than in the VMAT plan (7.65 Gy/10.88 Gy, p < 0.001 ), the Dice Similarity Coefficient (DSC) was found to be significantly better for the VMAT plans compared to the IMRT plans (0.77/0.58, resp., p value < 0.01 ), and an almost 50% reduction in the net treatment time was calculated for the VMAT compared to the IMRT plans (6.73 min/12.96 min, p < 0.001 ).

Conclusions.

In our report, VMAT provides better conformity, homogeneity, and safety profile.

The shorter treatment time is a major advantage and not only provides convenience to the painful patient but also contributes to the precision of this high dose radiation therapy.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Zach, Leor& Tsvang, Lev& Alezra, Dror& Ben Ayun, Maoz& Harel, Ran. 2016. Volumetric Modulated Arc Therapy for Spine Radiosurgery: Superior Treatment Planning and Delivery Compared to Static Beam Intensity Modulated Radiotherapy. BioMed Research International،Vol. 2016, no. 2016, pp.1-6.
https://search.emarefa.net/detail/BIM-1098567

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Zach, Leor…[et al.]. Volumetric Modulated Arc Therapy for Spine Radiosurgery: Superior Treatment Planning and Delivery Compared to Static Beam Intensity Modulated Radiotherapy. BioMed Research International No. 2016 (2016), pp.1-6.
https://search.emarefa.net/detail/BIM-1098567

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Zach, Leor& Tsvang, Lev& Alezra, Dror& Ben Ayun, Maoz& Harel, Ran. Volumetric Modulated Arc Therapy for Spine Radiosurgery: Superior Treatment Planning and Delivery Compared to Static Beam Intensity Modulated Radiotherapy. BioMed Research International. 2016. Vol. 2016, no. 2016, pp.1-6.
https://search.emarefa.net/detail/BIM-1098567

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1098567