Accuracy of Transcutaneous Carbon Dioxide Measurement in Premature Infants
المؤلفون المشاركون
Janaillac, Marie
Labarinas, Sonia
Pfister, Riccardo E.
Karam, Oliver
المصدر
Critical Care Research and Practice
العدد
المجلد 2016، العدد 2016 (31 ديسمبر/كانون الأول 2016)، ص ص. 1-5، 5ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2016-06-08
دولة النشر
مصر
عدد الصفحات
5
التخصصات الرئيسية
الملخص EN
Background.
In premature infants, maintaining blood partial pressure of carbon dioxide (pCO2) value within a narrow range is important to avoid cerebral lesions.
The aim of this study was to assess the accuracy of a noninvasive transcutaneous method (TcpCO2), compared to blood partial pressure of carbon dioxide (pCO2).
Methods.
Retrospective observational study in a tertiary neonatal intensive care unit.
We analyzed the correlation between blood pCO2 and transcutaneous values and the accuracy between the trends of blood pCO2 and TcpCO2 in all consecutive premature infants born at <33 weeks’ gestational age.
Results.
248 infants were included (median gestational age: 29 + 5 weeks and median birth weight: 1250 g), providing 1365 pairs of TcpCO2 and blood pCO2 values.
Pearson’s R correlation between these values was 0.58.
The mean bias was −0.93 kPa with a 95% confidence limit of agreement of −4.05 to +2.16 kPa.
Correlation between the trends of TcpCO2 and blood pCO2 values was good in only 39.6%.
Conclusions.
In premature infants, TcpCO2 was poorly correlated to blood pCO2, with a wide limit of agreement.
Furthermore, concordance between trends was equally low.
We warn about clinical decision-making on TcpCO2 alone when used as continuous monitoring.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Janaillac, Marie& Labarinas, Sonia& Pfister, Riccardo E.& Karam, Oliver. 2016. Accuracy of Transcutaneous Carbon Dioxide Measurement in Premature Infants. Critical Care Research and Practice،Vol. 2016, no. 2016, pp.1-5.
https://search.emarefa.net/detail/BIM-1099540
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Janaillac, Marie…[et al.]. Accuracy of Transcutaneous Carbon Dioxide Measurement in Premature Infants. Critical Care Research and Practice No. 2016 (2016), pp.1-5.
https://search.emarefa.net/detail/BIM-1099540
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Janaillac, Marie& Labarinas, Sonia& Pfister, Riccardo E.& Karam, Oliver. Accuracy of Transcutaneous Carbon Dioxide Measurement in Premature Infants. Critical Care Research and Practice. 2016. Vol. 2016, no. 2016, pp.1-5.
https://search.emarefa.net/detail/BIM-1099540
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-1099540
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر