Phantom Validation of Tc-99m Absolute Quantification in a SPECTCT Commercial Device

المؤلفون المشاركون

Gnesin, Silvano
Leite Ferreira, Paulo
Malterre, Jerome
Laub, Priscille
Verdun, Francis R.
Prior, John O.

المصدر

Computational and Mathematical Methods in Medicine

العدد

المجلد 2016، العدد 2016 (31 ديسمبر/كانون الأول 2016)، ص ص. 1-6، 6ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2016-12-14

دولة النشر

مصر

عدد الصفحات

6

التخصصات الرئيسية

الطب البشري

الملخص EN

Aim.

Similar to PET, absolute quantitative imaging is becoming available in commercial SPECT/CT devices.

This study’s goal was to assess quantitative accuracy of activity recovery as a function of image reconstruction parameters and count statistics in a variety of phantoms.

Materials and Methods.

We performed quantitative Tc99m-SPECT/CT acquisitions (Siemens Symbia Intevo, Erlangen, Germany) of a uniform cylindrical, NEMA/IEC, and an anthropomorphic abdominal phantom.

Background activity concentrations tested ranged: 2–80 kBq/mL.

SPECT acquisitions used 120 projections (20 s/projection).

Reconstructions were performed with the proprietary iterative conjugate gradient algorithm.

NEMA phantom reconstructions were obtained as a function of the iteration number (range: 4–48).

Recovery coefficients, hot contrast, relative lung error (NEMA phantom), and image noise were assessed.

Results.

In all cases, absolute activity and activity concentration were measured within 10% of the expected value.

Recovery coefficients and hot contrast in hot inserts did not vary appreciably with count statistics.

RC converged at 16 iterations for insert size > 22 mm.

Relative lung errors were comparable to PET levels indicating the efficient integration of attenuation and scatter corrections with adequate detector modeling.

Conclusions.

The tested device provided accurate activity recovery within 10% of correct values; these performances are comparable to current generation PET/CT systems.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Gnesin, Silvano& Leite Ferreira, Paulo& Malterre, Jerome& Laub, Priscille& Prior, John O.& Verdun, Francis R.. 2016. Phantom Validation of Tc-99m Absolute Quantification in a SPECTCT Commercial Device. Computational and Mathematical Methods in Medicine،Vol. 2016, no. 2016, pp.1-6.
https://search.emarefa.net/detail/BIM-1100131

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Gnesin, Silvano…[et al.]. Phantom Validation of Tc-99m Absolute Quantification in a SPECTCT Commercial Device. Computational and Mathematical Methods in Medicine No. 2016 (2016), pp.1-6.
https://search.emarefa.net/detail/BIM-1100131

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Gnesin, Silvano& Leite Ferreira, Paulo& Malterre, Jerome& Laub, Priscille& Prior, John O.& Verdun, Francis R.. Phantom Validation of Tc-99m Absolute Quantification in a SPECTCT Commercial Device. Computational and Mathematical Methods in Medicine. 2016. Vol. 2016, no. 2016, pp.1-6.
https://search.emarefa.net/detail/BIM-1100131

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1100131