Kinect-Based Correction of Overexposure Artifacts in Knee Imaging with C-Arm CT Systems
المؤلفون المشاركون
Fahrig, Rebecca
Rausch, Johannes
Choi, Jang-Hwan
Hinshaw, Waldo
Schebesch, Frank
Haase, Sven
Wasza, Jakob
Riess, Christian
Hornegger, J.
Maier, Andreas
المصدر
International Journal of Biomedical Imaging
العدد
المجلد 2016، العدد 2016 (31 ديسمبر/كانون الأول 2016)، ص ص. 1-15، 15ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2016-07-19
دولة النشر
مصر
عدد الصفحات
15
التخصصات الرئيسية
الملخص EN
Objective.
To demonstrate a novel approach of compensating overexposure artifacts in CT scans of the knees without attaching any supporting appliances to the patient.
C-Arm CT systems offer the opportunity to perform weight-bearing knee scans on standing patients to diagnose diseases like osteoarthritis.
However, one serious issue is overexposure of the detector in regions close to the patella, which can not be tackled with common techniques.
Methods.
A Kinect camera is used to algorithmically remove overexposure artifacts close to the knee surface.
Overexposed near-surface knee regions are corrected by extrapolating the absorption values from more reliable projection data.
To achieve this, we develop a cross-calibration procedure to transform surface points from the Kinect to CT voxel coordinates.
Results.
Artifacts at both knee phantoms are reduced significantly in the reconstructed data and a major part of the truncated regions is restored.
Conclusion.
The results emphasize the feasibility of the proposed approach.
The accuracy of the cross-calibration procedure can be increased to further improve correction results.
Significance.
The correction method can be extended to a multi-Kinect setup for use in real-world scenarios.
Using depth cameras does not require prior scans and offers the possibility of a temporally synchronized correction of overexposure artifacts.
To achieve this, we develop a cross-calibration procedure to transform surface points from the Kinect to CT voxel coordinates.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Rausch, Johannes& Maier, Andreas& Fahrig, Rebecca& Choi, Jang-Hwan& Hinshaw, Waldo& Schebesch, Frank…[et al.]. 2016. Kinect-Based Correction of Overexposure Artifacts in Knee Imaging with C-Arm CT Systems. International Journal of Biomedical Imaging،Vol. 2016, no. 2016, pp.1-15.
https://search.emarefa.net/detail/BIM-1105408
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Rausch, Johannes…[et al.]. Kinect-Based Correction of Overexposure Artifacts in Knee Imaging with C-Arm CT Systems. International Journal of Biomedical Imaging No. 2016 (2016), pp.1-15.
https://search.emarefa.net/detail/BIM-1105408
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Rausch, Johannes& Maier, Andreas& Fahrig, Rebecca& Choi, Jang-Hwan& Hinshaw, Waldo& Schebesch, Frank…[et al.]. Kinect-Based Correction of Overexposure Artifacts in Knee Imaging with C-Arm CT Systems. International Journal of Biomedical Imaging. 2016. Vol. 2016, no. 2016, pp.1-15.
https://search.emarefa.net/detail/BIM-1105408
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-1105408
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
![](/images/ebook-kashef.png)
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر
![](/images/kashef-image.png)