The Reliability of a Novel Automated System for ANA Immunofluorescence Analysis in Daily Clinical Practice

المؤلفون المشاركون

Alsuwaidi, Mohammed
Dollinger, Margit
Ehrenstein, Boris
Fleck, Martin

المصدر

International Journal of Rheumatology

العدد

المجلد 2016، العدد 2016 (31 ديسمبر/كانون الأول 2016)، ص ص. 1-5، 5ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2016-05-09

دولة النشر

مصر

عدد الصفحات

5

التخصصات الرئيسية

الطب البشري

الملخص EN

Automated interpretation (AI) systems for antinuclear antibody (ANA) analysis have been introduced based on assessment of indirect immunofluorescence (IIF) patterns.

The diagnostic performance of a novel automated IIF reading system was compared with visual interpretation (VI) of IIF in daily clinical practice to evaluate the reduction of workload.

ANA-IIF tests of consecutive serum samples from patients with suspected connective tissue disease were carried out using HEp-2 cells according to routine clinical care.

AI was performed using a visual analyser (Zenit G-Sight, Menarini, Germany).

Agreement rates between ANA results by AI and VI were calculated.

Of the 336 samples investigated, VI yielded 205 (61%) negative, 42 (13%) ambiguous, and 89 (26%) positive results, whereas 82 (24%) were determined to be negative, 176 (52%) ambiguous, and 78 (24%) positive by AI.

AI displayed a diagnostic accuracy of 175/336 samples (52%) with a kappa coefficient of 0.34 compared to VI being the gold standard.

Solely relying on AI, with VI only performed for all ambiguous samples by AI, would have missed 1 of 89 (1%) positive results by VI and misclassified 2 of 205 (1%) negative results by VI as positive.

The use of AI in daily clinical practice resulted only in a moderate reduction of the VI workload (82 of 336 samples: 24%).

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Alsuwaidi, Mohammed& Dollinger, Margit& Fleck, Martin& Ehrenstein, Boris. 2016. The Reliability of a Novel Automated System for ANA Immunofluorescence Analysis in Daily Clinical Practice. International Journal of Rheumatology،Vol. 2016, no. 2016, pp.1-5.
https://search.emarefa.net/detail/BIM-1106908

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Alsuwaidi, Mohammed…[et al.]. The Reliability of a Novel Automated System for ANA Immunofluorescence Analysis in Daily Clinical Practice. International Journal of Rheumatology No. 2016 (2016), pp.1-5.
https://search.emarefa.net/detail/BIM-1106908

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Alsuwaidi, Mohammed& Dollinger, Margit& Fleck, Martin& Ehrenstein, Boris. The Reliability of a Novel Automated System for ANA Immunofluorescence Analysis in Daily Clinical Practice. International Journal of Rheumatology. 2016. Vol. 2016, no. 2016, pp.1-5.
https://search.emarefa.net/detail/BIM-1106908

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1106908