Comparison between a New Optical Biometry Device and an Anterior Segment Optical Coherence Tomographer for Measuring Central Corneal Thickness and Anterior Chamber Depth

المؤلفون المشاركون

Huang, Jinhai
Lu, Weicong
Wang, Qinmei
Wang, Chengfang
Bao, Fang-Jun
Chen, Hao
Yu, Xinxin
Savini, G.

المصدر

Journal of Ophthalmology

العدد

المجلد 2016، العدد 2016 (31 ديسمبر/كانون الأول 2016)، ص ص. 1-5، 5ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2016-06-14

دولة النشر

مصر

عدد الصفحات

5

التخصصات الرئيسية

الطب البشري

الملخص EN

Purpose.

To compare between a new optical biometer (AL-Scan, Nidek Co., Aichi, Japan) and an anterior segment optical coherence tomographer (Visante AS-OCT, Carl Zeiss Meditec, Dublin, USA) for measuring central corneal thickness (CCT), anterior chamber depth (ACD), and aqueous depth (AD).

Methods.

Sixty-three eyes of 63 normal subjects were examined with AL-Scan and Visante AS-OCT in this prospective study.

One eye per subject was measured three times with both devices to record their CCT, ACD, and AD.

All procedures were performed by the same operator.

Agreement between the two devices was assessed using paired t-tests, Bland-Altman plots, and 95% limits of agreement (LoA).

Results.

The mean CCT, ACD, and AD measured by AL-Scan were 538.59 ± 27.37 μm, 3.70 ± 0.30 mm, and 3.16 ± 0.30 mm, respectively.

The mean values obtained by the Visante OCT were 536.14 ± 26.61 μm for CCT, 3.71 ± 0.29 mm for ACD, and 3.17 ± 0.29 mm for AD.

The mean CCT by the AL-Scan was higher than that obtained by the Visante AS-OCT (difference = 2.45 ± 6.07 μm, P < 0.05 ).

The differences in ACD and AD measurements were not statistically significant.

The 95% LoA of CCT, ACD, and AD were between −9.44 and 14.35 μm, −0.15 and 0.12 mm, and −0.15 and 0.12 mm, respectively.

Conclusions.

Since these two devices were comparable for measuring CCT, ACD, and AD, their results can be interchangeably used in the clinic.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Huang, Jinhai& Lu, Weicong& Savini, G.& Chen, Hao& Wang, Chengfang& Yu, Xinxin…[et al.]. 2016. Comparison between a New Optical Biometry Device and an Anterior Segment Optical Coherence Tomographer for Measuring Central Corneal Thickness and Anterior Chamber Depth. Journal of Ophthalmology،Vol. 2016, no. 2016, pp.1-5.
https://search.emarefa.net/detail/BIM-1110024

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Huang, Jinhai…[et al.]. Comparison between a New Optical Biometry Device and an Anterior Segment Optical Coherence Tomographer for Measuring Central Corneal Thickness and Anterior Chamber Depth. Journal of Ophthalmology No. 2016 (2016), pp.1-5.
https://search.emarefa.net/detail/BIM-1110024

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Huang, Jinhai& Lu, Weicong& Savini, G.& Chen, Hao& Wang, Chengfang& Yu, Xinxin…[et al.]. Comparison between a New Optical Biometry Device and an Anterior Segment Optical Coherence Tomographer for Measuring Central Corneal Thickness and Anterior Chamber Depth. Journal of Ophthalmology. 2016. Vol. 2016, no. 2016, pp.1-5.
https://search.emarefa.net/detail/BIM-1110024

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1110024