Biomechanical Measurement of Rabbit Cornea by a Modified Scheimpflug Device
المؤلفون المشاركون
Zheng, Danying
Zhang, Bo
Gu, Jianjun
Zhang, Xiaoxiao
Yang, Bin
Wang, Zheng
المصدر
العدد
المجلد 2016، العدد 2016 (31 ديسمبر/كانون الأول 2016)، ص ص. 1-6، 6ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2016-07-03
دولة النشر
مصر
عدد الصفحات
6
التخصصات الرئيسية
الملخص EN
Purpose.
To explore the probability and variation in biomechanical measurements of rabbit cornea by a modified Scheimpflug device.
Methods.
A modified Scheimpflug device was developed by imaging anterior segment of the model imitating the intact eye at various posterior pressures.
The eight isolated rabbit corneas were mounted on the Barron artificial chamber and images of the anterior segment were taken at posterior pressures of 15, 30, 45, 60, and 75 mmHg by the device.
The repeatability and reliability of the parameters including CCT, ACD, ACV, and CV were evaluated at each posterior pressure.
All the variations of the parameters at the different posterior pressures were calculated.
Results.
All parameters showed good intraobserver reliability (Cronbach’s alpha; intraclass correlation coefficient, α, ICC > 0.96) and repeatability in the modified Scheimpflug device.
With the increase of posterior pressures, the ratio of CCT decreased linearly and the bulk modulus gradually reduced to a platform.
The increase of ACD was almost linear with the posterior pressures elevated.
Conclusions.
The modified Scheimpflug device was a valuable tool to investigate the biomechanics of the cornea.
The posterior pressure 15–75 mmHg range produced small viscoelastic deformations and nearly linear pressure-deformation response in the rabbit cornea.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Zhang, Bo& Gu, Jianjun& Zhang, Xiaoxiao& Yang, Bin& Wang, Zheng& Zheng, Danying. 2016. Biomechanical Measurement of Rabbit Cornea by a Modified Scheimpflug Device. Journal of Ophthalmology،Vol. 2016, no. 2016, pp.1-6.
https://search.emarefa.net/detail/BIM-1110083
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Zhang, Bo…[et al.]. Biomechanical Measurement of Rabbit Cornea by a Modified Scheimpflug Device. Journal of Ophthalmology No. 2016 (2016), pp.1-6.
https://search.emarefa.net/detail/BIM-1110083
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Zhang, Bo& Gu, Jianjun& Zhang, Xiaoxiao& Yang, Bin& Wang, Zheng& Zheng, Danying. Biomechanical Measurement of Rabbit Cornea by a Modified Scheimpflug Device. Journal of Ophthalmology. 2016. Vol. 2016, no. 2016, pp.1-6.
https://search.emarefa.net/detail/BIM-1110083
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-1110083
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر