Measurement and Analysis of Gait Pattern during Stair Walk for Improvement of Robotic Locomotion Rehabilitation System

المؤلفون المشاركون

Park, Sang-Eun
Ho, Ye-Ji
Chun, Min Ho
Choi, Jaesoon
Moon, Youngjin

المصدر

Applied Bionics and Biomechanics

العدد

المجلد 2019، العدد 2019 (31 ديسمبر/كانون الأول 2019)، ص ص. 1-12، 12ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2019-10-14

دولة النشر

مصر

عدد الصفحات

12

التخصصات الرئيسية

الأحياء

الملخص EN

Background.

Robotic locomotion rehabilitation systems have been used for gait training in patients who have had a stroke.

Most commercialized systems allow patients to perform simple exercises such as balancing or level walking, but an additional function such as stair-walk training is required to provide a wide range of recovery cycle rehabilitation.

In this study, we analyzed stair-gait patterns and applied the result to a robotic rehabilitation system that can provide a vertical motion of footplates.

Methods.

To obtain applicable data for the robotic system with vertically movable footplates, stair-walk action was measured using an optical marker-based motion capture system.

The spatial position data of joints during stair walking was obtained from six healthy adults who participated in the experiment.

The measured marker data were converted into joint kinematic data by using an algorithm that included resampling and normalization.

The spatial position data are represented as angular trajectories and the relative displacement of each joint on the anatomical sagittal plane and movements of hip joints on the anatomical transverse plane.

Results.

The average range of motion (ROM) of each joint was estimated as (−6.75°,48.69°) at the hip, 8.20°,93.78° at the knee, and −17.78°,11.75° at the ankle during ascent and as 6.41°,31.67° at the hip, 7.38°,91.93° at the knee, and −24.89°,24.18° at the ankle during descent.

Additionally, we attempted to create a more natural stair-gait pattern by analyzing the movement of the hip on the anatomical transverse plane.

The hip movements were estimated to within ±1.57 cm and ±2.00 cm for hip translation and to within ±2.52° and ±2.70° for hip rotation during stair ascent and stair descent, respectively.

Conclusions.

Based on the results, standard patterns of stair ascent and stair descent were derived and applied to a lower-limb rehabilitation robot with vertically movable footplates.

The relative trajectory from the experiment ascertained that the function of stair walking in the robotic system properly worked within a normal ROM.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Park, Sang-Eun& Ho, Ye-Ji& Chun, Min Ho& Choi, Jaesoon& Moon, Youngjin. 2019. Measurement and Analysis of Gait Pattern during Stair Walk for Improvement of Robotic Locomotion Rehabilitation System. Applied Bionics and Biomechanics،Vol. 2019, no. 2019, pp.1-12.
https://search.emarefa.net/detail/BIM-1114564

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Park, Sang-Eun…[et al.]. Measurement and Analysis of Gait Pattern during Stair Walk for Improvement of Robotic Locomotion Rehabilitation System. Applied Bionics and Biomechanics No. 2019 (2019), pp.1-12.
https://search.emarefa.net/detail/BIM-1114564

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Park, Sang-Eun& Ho, Ye-Ji& Chun, Min Ho& Choi, Jaesoon& Moon, Youngjin. Measurement and Analysis of Gait Pattern during Stair Walk for Improvement of Robotic Locomotion Rehabilitation System. Applied Bionics and Biomechanics. 2019. Vol. 2019, no. 2019, pp.1-12.
https://search.emarefa.net/detail/BIM-1114564

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1114564