Background and Objective.
A common reconstruction procedure after a wide resection of bone tumors around the knee is endoprosthetic knee replacement.
The aim of this study was to investigate the characteristics of bone injury of the patient after endoprosthetic knee replacement during walking.
Methods.
A subject-specific finite element model of the femur-prosthesis-tibia complex was established via CT scans.
To obtain its physiologically realistic loading environments, the musculoskeletal inverse dynamic analysis was implemented.
The extracted muscle forces and ground forces were then applied to the finite element model to investigate bone stress distribution at various stages of the gait cycle.
Results.
The maximum femur stress of each stage varied from 33.14 MPa to 70.61 MPa in the gait cycle.
The stress concentration position with a distance of 267.2 mm to the tibial plateau showed a good agreement with the patient injury data.
Conclusions.
Overall results indicated the reasonability of the simulation method to determine loading environments and injury characteristics which the patient experienced with knee endoprosthesis during walking.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Mo, Fuhao& Zhang, Haotian& Zhao, Siqi& Xiao, Zhi& Liu, Tang. 2019. Coupling Musculoskeletal Dynamics and Subject-Specific Finite Element Analysis of Femoral Cortical Bone Failure after Endoprosthetic Knee Replacement. Applied Bionics and Biomechanics،Vol. 2019, no. 2019, pp.1-8.
https://search.emarefa.net/detail/BIM-1114618
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Mo, Fuhao…[et al.]. Coupling Musculoskeletal Dynamics and Subject-Specific Finite Element Analysis of Femoral Cortical Bone Failure after Endoprosthetic Knee Replacement. Applied Bionics and Biomechanics No. 2019 (2019), pp.1-8.
https://search.emarefa.net/detail/BIM-1114618
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Mo, Fuhao& Zhang, Haotian& Zhao, Siqi& Xiao, Zhi& Liu, Tang. Coupling Musculoskeletal Dynamics and Subject-Specific Finite Element Analysis of Femoral Cortical Bone Failure after Endoprosthetic Knee Replacement. Applied Bionics and Biomechanics. 2019. Vol. 2019, no. 2019, pp.1-8.
https://search.emarefa.net/detail/BIM-1114618
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-1114618
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر