Development of a Novel Micro Photoionization Detector for Rapid Volatile Organic Compounds Measurement
المؤلفون المشاركون
Ma, Xu
Zhou, Qi
Zhang, Sixiang
Zhang, Xu
Zhou, Wei
المصدر
Applied Bionics and Biomechanics
العدد
المجلد 2018، العدد 2018 (31 ديسمبر/كانون الأول 2018)، ص ص. 1-9، 9ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2018-09-05
دولة النشر
مصر
عدد الصفحات
9
التخصصات الرئيسية
الملخص EN
The simulation of the gas flow field and electrostatic field in the photoionization detector by COMSOL was conducted based on principle investigation in the present study.
Under the guidance of simulation results, structural optimization was carried out to significantly reduce the dead volume of the ionization chamber, and finally, the relationship between offset voltage and collection efficiency was obtained which led to a remarkable increase in the collection efficiency of charged ions in the photoionization detector.
Then an ionization chamber with low interference and fast response was developed.
Then experiment was performed with toluene as a VOCs gas under the condition of optimal gas flow rate of 50 ml, UV lamp ionization energy of 10.86 eV.
The results showed that the ion collection efficiency reached 91% at a bias voltage of 150 V.
Moreover, a preferred linearity of 99.99% was obtained, and a ppb level of LOD can be achieved.
The determination results well-fitted the relationship between offset voltage and the response value obtained in the simulation.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Zhou, Qi& Zhang, Sixiang& Zhang, Xu& Ma, Xu& Zhou, Wei. 2018. Development of a Novel Micro Photoionization Detector for Rapid Volatile Organic Compounds Measurement. Applied Bionics and Biomechanics،Vol. 2018, no. 2018, pp.1-9.
https://search.emarefa.net/detail/BIM-1114681
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Zhou, Qi…[et al.]. Development of a Novel Micro Photoionization Detector for Rapid Volatile Organic Compounds Measurement. Applied Bionics and Biomechanics No. 2018 (2018), pp.1-9.
https://search.emarefa.net/detail/BIM-1114681
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Zhou, Qi& Zhang, Sixiang& Zhang, Xu& Ma, Xu& Zhou, Wei. Development of a Novel Micro Photoionization Detector for Rapid Volatile Organic Compounds Measurement. Applied Bionics and Biomechanics. 2018. Vol. 2018, no. 2018, pp.1-9.
https://search.emarefa.net/detail/BIM-1114681
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-1114681
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر