Mechanical Testing of a Novel Fastening Device to Improve Scoliosis Bracing Biomechanics for Treating Adolescent Idiopathic Scoliosis
المؤلفون المشاركون
Di’Angelo, Denis
Chung, Chloe L.
Kelly, Derek M.
Sawyer, Jeffery R.
Steele, Jack R.
Tate, Terry S.
Bateman, Cody K.
المصدر
Applied Bionics and Biomechanics
العدد
المجلد 2018، العدد 2018 (31 ديسمبر/كانون الأول 2018)، ص ص. 1-10، 10ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2018-08-12
دولة النشر
مصر
عدد الصفحات
10
التخصصات الرئيسية
الملخص EN
Velcro fastening straps are commonly used to secure a scoliosis brace around the upper body and apply corrective forces to the spine.
However, strap loosening and tension loss have been reported that reduce spinal correction and treatment efficacy.
A novel fastening device, or controlled tension unit (CTU), was designed to overcome these limitations.
A scoliosis analog model (SAM) was used to biomechanically compare the CTU fasteners and posterior Velcro straps on a conventional brace (CB) as well as on a modified brace (MB) that included a dynamic cantilever apical pad section.
Brace configurations tested were (1) CB with posterior Velcro straps, (2) CB with posterior CTU fasteners, (3) MB with posterior Velcro straps, and (4) MB with posterior CTU fasteners.
MB configurations were tested with 0 N, 35.6 N, and 71.2 N CTU fasteners applied across the apical pad flap.
Three-dimensional forces and moments were measured at both ends of the SAM.
The CTU fasteners provided the same corrective spinal loads as Velcro straps when tensioned to the same level on the CB configuration and can be used as an alternative fastening system.
Dynamically loading the apical flap increased the distractive forces applied to the spine without affecting tension in the fastening straps.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Chung, Chloe L.& Kelly, Derek M.& Sawyer, Jeffery R.& Steele, Jack R.& Tate, Terry S.& Bateman, Cody K.…[et al.]. 2018. Mechanical Testing of a Novel Fastening Device to Improve Scoliosis Bracing Biomechanics for Treating Adolescent Idiopathic Scoliosis. Applied Bionics and Biomechanics،Vol. 2018, no. 2018, pp.1-10.
https://search.emarefa.net/detail/BIM-1114755
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Chung, Chloe L.…[et al.]. Mechanical Testing of a Novel Fastening Device to Improve Scoliosis Bracing Biomechanics for Treating Adolescent Idiopathic Scoliosis. Applied Bionics and Biomechanics No. 2018 (2018), pp.1-10.
https://search.emarefa.net/detail/BIM-1114755
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Chung, Chloe L.& Kelly, Derek M.& Sawyer, Jeffery R.& Steele, Jack R.& Tate, Terry S.& Bateman, Cody K.…[et al.]. Mechanical Testing of a Novel Fastening Device to Improve Scoliosis Bracing Biomechanics for Treating Adolescent Idiopathic Scoliosis. Applied Bionics and Biomechanics. 2018. Vol. 2018, no. 2018, pp.1-10.
https://search.emarefa.net/detail/BIM-1114755
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-1114755
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر