Comparative Evaluation of a New Depth of Anesthesia Index in ConView® System and the Bispectral Index during Total Intravenous Anesthesia: A Multicenter Clinical Trial

المؤلفون المشاركون

Deng, Xiaoming
Xie, Keliang
Fu, Yang
Xu, Tao
Wei, Wei
Gao, Ping
Nie, Huang
Wang, Guolin
Tian, Ming
Yan, Min
Dong, Hailong
Yue, Yun

المصدر

BioMed Research International

العدد

المجلد 2019، العدد 2019 (31 ديسمبر/كانون الأول 2019)، ص ص. 1-7، 7ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2019-03-04

دولة النشر

مصر

عدد الصفحات

7

التخصصات الرئيسية

الطب البشري

الملخص EN

The performance of a new monitor for the depth of anesthesia (DOA), the Depth of Anesthesia Index (Ai) based on sample entropy (SampEn), 95% spectral edge frequency (95%SEF), and burst suppression ratio (BSR) was evaluated compared to Bispectral Index (BIS) during total intravenous anesthesia (TIVA).

144 patients in six medical centers were enrolled.

General anesthesia was induced with stepwise-increased target-controlled infusion (TCI) of propofol until loss of consciousness (LOC).

During surgery propofol was titrated according to BIS.

Both Ai and BIS were recorded.

Primary outcomes: the limits of agreement between Ai and BIS were -17.68 and 16.49, which were, respectively, -30.0% and 28.0% of the mean value of BIS.

Secondary outcomes: prediction probability (Pk) of BIS and Ai was 0.943 and 0.935 (p=0.102) during LOC and 0.928 and 0.918 (p=0.037) during recovery of consciousness (ROC).

And the values of BIS and Ai were 68.19 and 66.44 at 50%LOC, and 76.65 and 78.60 at 50%ROC.

A decrease or an increase of Ai was significantly greater than that of BIS when consciousness changes (during LOC: -9.13±10.20 versus -5.83±9.63, p<0.001; during ROC: 10.88±11.51 versus 5.32±7.53, p<0.001).

The conclusion is that Ai has similar characteristic of BIS as a DOA monitor and revealed the advantage of SampEn for indicating conscious level.

This trial is registered at Chinese Clinical Trial Registry with ChiCTR-IOR-16009471.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Fu, Yang& Xu, Tao& Xie, Keliang& Wei, Wei& Gao, Ping& Nie, Huang…[et al.]. 2019. Comparative Evaluation of a New Depth of Anesthesia Index in ConView® System and the Bispectral Index during Total Intravenous Anesthesia: A Multicenter Clinical Trial. BioMed Research International،Vol. 2019, no. 2019, pp.1-7.
https://search.emarefa.net/detail/BIM-1123103

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Fu, Yang…[et al.]. Comparative Evaluation of a New Depth of Anesthesia Index in ConView® System and the Bispectral Index during Total Intravenous Anesthesia: A Multicenter Clinical Trial. BioMed Research International No. 2019 (2019), pp.1-7.
https://search.emarefa.net/detail/BIM-1123103

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Fu, Yang& Xu, Tao& Xie, Keliang& Wei, Wei& Gao, Ping& Nie, Huang…[et al.]. Comparative Evaluation of a New Depth of Anesthesia Index in ConView® System and the Bispectral Index during Total Intravenous Anesthesia: A Multicenter Clinical Trial. BioMed Research International. 2019. Vol. 2019, no. 2019, pp.1-7.
https://search.emarefa.net/detail/BIM-1123103

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1123103