Optimum Configuration of Cannulated Compression Screws for the Fixation of Unstable Femoral Neck Fractures: Finite Element Analysis Evaluation

المؤلفون المشاركون

Li, Jiantao
Zhang, Hao
Li, Chen
Hao, Ming
Wang, Menglin
Zhou, Jianfeng
Han, Lin
Li, Lianting

المصدر

BioMed Research International

العدد

المجلد 2018، العدد 2018 (31 ديسمبر/كانون الأول 2018)، ص ص. 1-10، 10ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2018-12-09

دولة النشر

مصر

عدد الصفحات

10

التخصصات الرئيسية

الطب البشري

الملخص EN

Objectives.

In the present study, we evaluated the mechanical outcome of different configurations of cannulated compression screws for the fixation of Pauwels type III femoral neck fracture and the stress distribution around the holes corresponding to fixation protocol after screws removal.

Methods.

The Pauwels type III of femoral neck fracture was created in 3-matic software and the models of cannulated compression screws were constructed using UG-NX software.

Five fixation systems were assembled to the fracture models.

Abaqus software was used to perform the process of finite element analysis.

Values of stress distribution, maximum stress, model principal strains of proximal fragment, and stress distribution around the holes of femur model were recorded.

Results.

Stress of cannulated compression screws was intensely focused on the middle area of the screw near the fragment of each group.

Inverted triangle model showed the highest peak stress on screws under different phases of load.

Each screw dispersed some stresses, but at least one underwent the peak stress.

Fracture model fixed by inverted triangle configuration showed the lowest volume of yielding strain in the proximal fragment.

The area of higher stress around the holes was largest after triangle screws removal when compared with other four models.

Conclusions.

Our study indicated that different cannulated compression screws fixation configurations for the unstable femoral neck fractures showed the different mechanical efficiency.

Inverted triangular configuration showed the mechanical advantage and being less likely to cutout.

The fixation strategy of triangle configuration was least recommended if patients tended to remove the implants.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Li, Jiantao& Wang, Menglin& Zhou, Jianfeng& Han, Lin& Zhang, Hao& Li, Chen…[et al.]. 2018. Optimum Configuration of Cannulated Compression Screws for the Fixation of Unstable Femoral Neck Fractures: Finite Element Analysis Evaluation. BioMed Research International،Vol. 2018, no. 2018, pp.1-10.
https://search.emarefa.net/detail/BIM-1124208

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Li, Jiantao…[et al.]. Optimum Configuration of Cannulated Compression Screws for the Fixation of Unstable Femoral Neck Fractures: Finite Element Analysis Evaluation. BioMed Research International No. 2018 (2018), pp.1-10.
https://search.emarefa.net/detail/BIM-1124208

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Li, Jiantao& Wang, Menglin& Zhou, Jianfeng& Han, Lin& Zhang, Hao& Li, Chen…[et al.]. Optimum Configuration of Cannulated Compression Screws for the Fixation of Unstable Femoral Neck Fractures: Finite Element Analysis Evaluation. BioMed Research International. 2018. Vol. 2018, no. 2018, pp.1-10.
https://search.emarefa.net/detail/BIM-1124208

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1124208