![](/images/graphics-bg.png)
A Comparative Analysis of Methods for Evaluation of ECG Signal Quality after Compression
المؤلفون المشاركون
Němcová, Andrea
Smíšek, Radovan
Maršánová, Lucie
Smital, Lukáš
Vítek, Martin
المصدر
العدد
المجلد 2018، العدد 2018 (31 ديسمبر/كانون الأول 2018)، ص ص. 1-26، 26ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2018-07-18
دولة النشر
مصر
عدد الصفحات
26
التخصصات الرئيسية
الملخص EN
The assessment of ECG signal quality after compression is an essential part of the compression process.
Compression facilitates the signal archiving, speeds up signal transmission, and reduces the energy consumption.
Conversely, lossy compression distorts the signals.
Therefore, it is necessary to express the compression performance through both compression efficiency and signal quality.
This paper provides an overview of objective algorithms for the assessment of both ECG signal quality after compression and compression efficiency.
In this area, there is a lack of standardization, and there is no extensive review as such.
40 methods were tested in terms of their suitability for quality assessment.
For this purpose, the whole CSE database was used.
The tested signals were compressed using an algorithm based on SPIHT with varying efficiency.
As a reference, compressed signals were manually assessed by two experts and classified into three quality groups.
Owing to the experts’ classification, we determined corresponding ranges of selected quality evaluation methods’ values.
The suitability of the methods for quality assessment was evaluated based on five criteria.
For the assessment of ECG signal quality after compression, we recommend using a combination of these methods: PSim SDNN, QS, SNR1, MSE, PRDN1, MAX, STDERR, and WEDD SWT.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Němcová, Andrea& Smíšek, Radovan& Maršánová, Lucie& Smital, Lukáš& Vítek, Martin. 2018. A Comparative Analysis of Methods for Evaluation of ECG Signal Quality after Compression. BioMed Research International،Vol. 2018, no. 2018, pp.1-26.
https://search.emarefa.net/detail/BIM-1124712
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Němcová, Andrea…[et al.]. A Comparative Analysis of Methods for Evaluation of ECG Signal Quality after Compression. BioMed Research International No. 2018 (2018), pp.1-26.
https://search.emarefa.net/detail/BIM-1124712
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Němcová, Andrea& Smíšek, Radovan& Maršánová, Lucie& Smital, Lukáš& Vítek, Martin. A Comparative Analysis of Methods for Evaluation of ECG Signal Quality after Compression. BioMed Research International. 2018. Vol. 2018, no. 2018, pp.1-26.
https://search.emarefa.net/detail/BIM-1124712
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references