![](/images/graphics-bg.png)
Using a Somatosensory Controller to Assess Body Size for Size-Specific Dose Estimates in Computed Tomography
المؤلفون المشاركون
Han, Ruo-Ping
Liu, Yan-Lin
Wu, Jay
المصدر
العدد
المجلد 2018، العدد 2018 (31 ديسمبر/كانون الأول 2018)، ص ص. 1-7، 7ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2018-05-31
دولة النشر
مصر
عدد الصفحات
7
التخصصات الرئيسية
الملخص EN
Computed tomography (CT) has been widely used in the healthcare environment.
Presently, the radiation dose in CT is determined using the size-specific dose estimate (SSDE).
Accurate assessment of individual’s body size is essential for dose estimation.
In this study, we integrated a somatosensory controller with a CT scanner to measure patient’s anterior-posterior diameter (APD) and lateral diameter (LATD) and calculate the corresponding effective diameter (ED).
A total of 108 individuals with an average age of 38.6 years were enrolled in this study.
Microsoft Kinect was used to acquire the depth image of subjects.
A grayscale-to-surface height conversion curve was created using acrylic sheets for APD estimation.
The APD, LATD, and ED were measured and compared with the results obtained using F ruler and CT images.
The mean absolute differences for APD, LATD, and ED between Kinect and F ruler measurements were 5.2%, 1.3%, and 2.5%, respectively, while those between Kinect and CT measurements were 8.8%, 2.6%, and 5.0%, respectively.
Kinect can replace CT or F ruler for real-time body size measurements.
The use of the somatosensory controller has the advantages of simple, low cost, no radiation, and automatic calculation.
It can accurately estimate patient’s APD, LATD, and ED for SSDE.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Wu, Jay& Han, Ruo-Ping& Liu, Yan-Lin. 2018. Using a Somatosensory Controller to Assess Body Size for Size-Specific Dose Estimates in Computed Tomography. BioMed Research International،Vol. 2018, no. 2018, pp.1-7.
https://search.emarefa.net/detail/BIM-1125284
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Wu, Jay…[et al.]. Using a Somatosensory Controller to Assess Body Size for Size-Specific Dose Estimates in Computed Tomography. BioMed Research International No. 2018 (2018), pp.1-7.
https://search.emarefa.net/detail/BIM-1125284
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Wu, Jay& Han, Ruo-Ping& Liu, Yan-Lin. Using a Somatosensory Controller to Assess Body Size for Size-Specific Dose Estimates in Computed Tomography. BioMed Research International. 2018. Vol. 2018, no. 2018, pp.1-7.
https://search.emarefa.net/detail/BIM-1125284
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references