Image Quality and Patient-Specific Organ Doses in Stone Protocol CT: A Comparison of Traditional CT to Low Dose CT with Iterative Reconstruction

المؤلفون المشاركون

Canales, Benjamin K.
Pai, Raghav
Modh, Rishi
Lamoureux, Rebecca H.
Deitte, Lori
Wymer, David C.
Mench, Anna
Lipnharski, Izabella
Henriksen, Carl
Arreola, Manuel

المصدر

BioMed Research International

العدد

المجلد 2018، العدد 2018 (31 ديسمبر/كانون الأول 2018)، ص ص. 1-6، 6ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2018-09-27

دولة النشر

مصر

عدد الصفحات

6

التخصصات الرئيسية

الطب البشري

الملخص EN

Objective.

To compare organ specific radiation dose and image quality in kidney stone patients scanned with standard CT reconstructed with filtered back projection (FBP-CT) to those scanned with low dose CT reconstructed with iterative techniques (IR-CT).

Materials and Methods.

Over a one-year study period, adult kidney stone patients were retrospectively netted to capture the use of noncontrasted, stone protocol CT in one of six institutional scanners (four FBP and two IR).

To limit potential CT-unit use bias, scans were included only from days when all six scanners were functioning.

Organ dose was calculated using volumetric CT dose index and patient effective body diameter through validated conversion equations derived from previous cadaveric, dosimetry studies.

Board-certified radiologists, blinded to CT algorithm type, assessed stone characteristics, study noise, and image quality of both techniques.

Results.

FBP-CT (n=250) and IR-CT (n=90) groups were similar in regard to gender, race, body mass index (mean BMI = 30.3), and stone burden detected (mean size 5.4 ± 1.2 mm).

Mean organ-specific dose (OSD) was 54-62% lower across all organs for IR-CT compared to FBP-CT with particularly reduced doses (up to 4.6-fold) noted in patients with normal BMI range.

No differences were noted in radiological assessment of image quality or noise between the cohorts, and intrarater agreement was highly correlated for noise (AC2=0.873) and quality (AC2=0.874) between blinded radiologists.

Conclusions.

Image quality and stone burden assessment were maintained between standard FBP and low dose IR groups, but IR-CT decreased mean OSD by 50%.

Both urologists and radiologists should advocate for low dose CT, utilizing reconstructive protocols like IR, to reduce radiation exposure in their stone formers who undergo multiple CTs.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Pai, Raghav& Modh, Rishi& Lamoureux, Rebecca H.& Deitte, Lori& Wymer, David C.& Mench, Anna…[et al.]. 2018. Image Quality and Patient-Specific Organ Doses in Stone Protocol CT: A Comparison of Traditional CT to Low Dose CT with Iterative Reconstruction. BioMed Research International،Vol. 2018, no. 2018, pp.1-6.
https://search.emarefa.net/detail/BIM-1127100

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Pai, Raghav…[et al.]. Image Quality and Patient-Specific Organ Doses in Stone Protocol CT: A Comparison of Traditional CT to Low Dose CT with Iterative Reconstruction. BioMed Research International No. 2018 (2018), pp.1-6.
https://search.emarefa.net/detail/BIM-1127100

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Pai, Raghav& Modh, Rishi& Lamoureux, Rebecca H.& Deitte, Lori& Wymer, David C.& Mench, Anna…[et al.]. Image Quality and Patient-Specific Organ Doses in Stone Protocol CT: A Comparison of Traditional CT to Low Dose CT with Iterative Reconstruction. BioMed Research International. 2018. Vol. 2018, no. 2018, pp.1-6.
https://search.emarefa.net/detail/BIM-1127100

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1127100