The Dosimetric Impact of Shifts in Patient Positioning during Boron Neutron Capture Therapy for Brain Tumors

المؤلفون المشاركون

Hsu, S. M.
Chen, Yi-Wei
Chuang, Keh-Shih
Lee, Jia-Cheng
Hsu, Fang-Yuh
Chou, Fong-In
Yen, Sang-Hue
Wu, Yuan-Hung

المصدر

BioMed Research International

العدد

المجلد 2018، العدد 2018 (31 ديسمبر/كانون الأول 2018)، ص ص. 1-11، 11ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2018-10-01

دولة النشر

مصر

عدد الصفحات

11

التخصصات الرئيسية

الطب البشري

الملخص EN

Unlike conventional photon radiotherapy, sophisticated patient positioning tools are not available for boron neutron capture therapy (BNCT).

Thus, BNCT remains vulnerable to setup errors and intra-fractional patient motion.

The aim of this study was to estimate the impact of deviations in positioning on the dose administered by BNCT for brain tumors at the Tsing Hua open-pool reactor (THOR).

For these studies, a simulated head model was generated based on computed tomography (CT) images of a patient with a brain tumor.

A cylindrical brain tumor 3 cm in diameter and 5 cm in length was modeled at distances of 6.5 cm and 2.5 cm from the posterior scalp of this head model (T6.5 cm and T2.5 cm, respectively).

Radiation doses associated with positioning errors were evaluated for each distance, including left and right shifts, superior and inferior shifts, shifts from the central axis of the beam aperture, and outward shifts from the surface of the beam aperture.

Rotational and tilting effects were also evaluated.

The dose prescription was 20 Gray-equivalent (Gy-Eq) to 80 % of the tumor.

The treatment planning system, NCTPlan, was used to perform dose calculations.

The average decreases in mean tumor dose for T6.5 cm for the 1 cm, 2 cm, and 3 cm lateral shifts composed by left, right, superior, and inferior sides, were approximately 1 %, 6 %, and 11 %, respectively, compared to the dose administered to the initial tumor position.

The decreases in mean tumor dose for T6.5 cm were approximately 5 %, 11 %, and 15 % for the 1 cm, 2 cm, and 3 cm outward shifts, respectively.

For a superficial tumor at T2.5cm, no significant decrease in average mean tumor dose was observed following lateral shifts of 1 cm.

Rotational and tilting up to 15° did not result in significant difference to the tumor dose.

Dose differences to the normal tissues as a result of the shifts in positioning were also minimal.

Taken together, these data demonstrate that the mean dose administered to tumors at greater depths is potentially more vulnerable to deviations in positioning, and greater shift distances resulted in reduced mean tumor doses at the THOR.

Moreover, these data provide an estimation of dose differences that are caused by setup error or intra-fractional motion during BNCT, and these may facilitate more accurate predictions of actual patient dose in future treatments.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Lee, Jia-Cheng& Chen, Yi-Wei& Chuang, Keh-Shih& Hsu, Fang-Yuh& Chou, Fong-In& Hsu, S. M.…[et al.]. 2018. The Dosimetric Impact of Shifts in Patient Positioning during Boron Neutron Capture Therapy for Brain Tumors. BioMed Research International،Vol. 2018, no. 2018, pp.1-11.
https://search.emarefa.net/detail/BIM-1127553

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Lee, Jia-Cheng…[et al.]. The Dosimetric Impact of Shifts in Patient Positioning during Boron Neutron Capture Therapy for Brain Tumors. BioMed Research International No. 2018 (2018), pp.1-11.
https://search.emarefa.net/detail/BIM-1127553

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Lee, Jia-Cheng& Chen, Yi-Wei& Chuang, Keh-Shih& Hsu, Fang-Yuh& Chou, Fong-In& Hsu, S. M.…[et al.]. The Dosimetric Impact of Shifts in Patient Positioning during Boron Neutron Capture Therapy for Brain Tumors. BioMed Research International. 2018. Vol. 2018, no. 2018, pp.1-11.
https://search.emarefa.net/detail/BIM-1127553

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1127553