Reliability and Accuracy of Peri-Interventional Stenosis Grading in Peripheral Artery Disease Using Color-Coded Quantitative Fluoroscopy: A Phantom Study Comparing a Clinical and Scientific Postprocessing Software

المؤلفون المشاركون

Ghibes, Patrick
Partovi, Sasan
Grözinger, Gerd
Martirosian, Petros
Nikolaou, Konstantin
Ketelsen, Dominik
Grosse, Ulrich
Syha, Roland
Schick, Fritz

المصدر

BioMed Research International

العدد

المجلد 2018، العدد 2018 (31 ديسمبر/كانون الأول 2018)، ص ص. 1-11، 11ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2018-07-24

دولة النشر

مصر

عدد الصفحات

11

التخصصات الرئيسية

الطب البشري

الملخص EN

Purpose.

To assess quantitative stenosis grading by color-coded fluoroscopy using an in vitro pulsatile flow phantom.

Methods.

Three different stenotic tubes (80%, 60%, and 40% diameter restriction) and a nonstenotic reference tube were compared regarding their different flow behavior by using contrast-enhanced fluoroscopy with a flat-detector system for visualisation purposes.

Time-density curves (TDC), area under the curve (AUC), time-to-peak (TTP), and different ROI sizes were analyzed in three independent measurements using two different postprocessing software solutions.

In addition, exemplary TDCs of a patient with a high-grade stenosis before and after stent angioplasty were acquired.

Results.

Color-coded fluoroscopy enabled depiction of differences in AUC and TDC between high-grade (80%), middle (60%), low-grade (40%), and nonstenotic tubes.

The best correlation between high-, middle-, and low-grade stenosis was appreciated in ROIs behind the stenosis.

This effect was enhanced by using longer integration times (5s, 7s) and a maximum frame rate of image acquisition for analysis (correlation coefficient rho=0.9284 at 5s).

TTP showed no significant differences between high- and low-grade stenosis.

Conclusions.

Various clinical studies in the literature already demonstrated reproducible and reliable stenosis grading by analyzing TDCs acquired with color-coded fluoroscopy.

In contrast to TTP, AUC values derived in ROIs behind the stenosis proved to be reliable parameters for stenosis grading.

However, our results also demonstrate that several factors are able to significantly impact the evaluation of AUC values.

More precisely, accuracy of acquired AUC values can be improved by choosing longer integration times, a large ROI size adapted to the vessel diameter, and a higher frame rate of image acquisition.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Ghibes, Patrick& Partovi, Sasan& Grözinger, Gerd& Martirosian, Petros& Schick, Fritz& Nikolaou, Konstantin…[et al.]. 2018. Reliability and Accuracy of Peri-Interventional Stenosis Grading in Peripheral Artery Disease Using Color-Coded Quantitative Fluoroscopy: A Phantom Study Comparing a Clinical and Scientific Postprocessing Software. BioMed Research International،Vol. 2018, no. 2018, pp.1-11.
https://search.emarefa.net/detail/BIM-1127749

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Ghibes, Patrick…[et al.]. Reliability and Accuracy of Peri-Interventional Stenosis Grading in Peripheral Artery Disease Using Color-Coded Quantitative Fluoroscopy: A Phantom Study Comparing a Clinical and Scientific Postprocessing Software. BioMed Research International No. 2018 (2018), pp.1-11.
https://search.emarefa.net/detail/BIM-1127749

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Ghibes, Patrick& Partovi, Sasan& Grözinger, Gerd& Martirosian, Petros& Schick, Fritz& Nikolaou, Konstantin…[et al.]. Reliability and Accuracy of Peri-Interventional Stenosis Grading in Peripheral Artery Disease Using Color-Coded Quantitative Fluoroscopy: A Phantom Study Comparing a Clinical and Scientific Postprocessing Software. BioMed Research International. 2018. Vol. 2018, no. 2018, pp.1-11.
https://search.emarefa.net/detail/BIM-1127749

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1127749