Deep Transcranial Magnetic Stimulation: Improved Coil Design and Assessment of the Induced Fields Using MIDA Model

المؤلفون المشاركون

Martens, L.
Samoudi, Amine M.
Tanghe, Emmeric
Joseph, Wout

المصدر

BioMed Research International

العدد

المجلد 2018، العدد 2018 (31 ديسمبر/كانون الأول 2018)، ص ص. 1-9، 9ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2018-06-05

دولة النشر

مصر

عدد الصفحات

9

التخصصات الرئيسية

الطب البشري

الملخص EN

Stimulation of deep brain structures by transcranial magnetic stimulation (TMS) is a method for activating deep neurons in the brain and can be beneficial for the treatment of psychiatric and neurological disorders.

To numerically investigate the possibility for deeper brain stimulation (electric fields reaching the hippocampus, the nucleus accumbens, and the cerebellum), combined TMS coils using the double-cone coil with the Halo coil (HDA) were modeled and investigated.

Numerical simulations were performed using MIDA: a new multimodal imaging-based detailed anatomical model of the human head and neck.

The 3D distributions of magnetic flux density and electric field were calculated.

The percentage of volume of each tissue that is exposed to electric field amplitude equal or greater than 50% of the maximum amplitude of E in the cortex for each coil was calculated to quantify the electric field spread (V50).

Results show that only the HDA coil can spread electric fields to the hippocampus, the nucleus accumbens, and the cerebellum with V50 equal to 0.04%, 1.21%, and 6.2%, respectively.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Samoudi, Amine M.& Tanghe, Emmeric& Martens, L.& Joseph, Wout. 2018. Deep Transcranial Magnetic Stimulation: Improved Coil Design and Assessment of the Induced Fields Using MIDA Model. BioMed Research International،Vol. 2018, no. 2018, pp.1-9.
https://search.emarefa.net/detail/BIM-1128244

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Samoudi, Amine M.…[et al.]. Deep Transcranial Magnetic Stimulation: Improved Coil Design and Assessment of the Induced Fields Using MIDA Model. BioMed Research International No. 2018 (2018), pp.1-9.
https://search.emarefa.net/detail/BIM-1128244

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Samoudi, Amine M.& Tanghe, Emmeric& Martens, L.& Joseph, Wout. Deep Transcranial Magnetic Stimulation: Improved Coil Design and Assessment of the Induced Fields Using MIDA Model. BioMed Research International. 2018. Vol. 2018, no. 2018, pp.1-9.
https://search.emarefa.net/detail/BIM-1128244

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1128244