Inherent Geometry Correction for Diffusion EPI Using the Reference Echoes as Navigators

المؤلفون المشاركون

van den Brink, Johan S.
Koonen, Jos J.

المصدر

Concepts in Magnetic Resonance :Part B : Magnetic Resonance Engineering.

العدد

المجلد 2019، العدد 2019 (31 ديسمبر/كانون الأول 2019)، ص ص. 1-8، 8ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2019-05-26

دولة النشر

مصر

عدد الصفحات

8

التخصصات الرئيسية

الفيزياء
الطب البشري

الملخص EN

Diffusion-weighted EPI has become an indispensable tool in body MRI.

Geometric distortions due to field inhomogeneities are more prominent at large field–of–view and require correction for comparison with T2W TSE.

Several known correction methods require acquisition of additional lengthy scans, which are difficult to apply in body imaging.

We implement and evaluate a geometry correction method based on the already available non phase-encoded EPI reference data used for Nyquist ghost removal.

The method is shown to provide accurate and robust global geometry correction in the absence of strong, local phase offsets.

It does not require additional time for calibrations and is directly compatible with parallel imaging methods.

The resulting images can serve as improved starting point for additional geometry correction methods relying on feature extraction and registration.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

van den Brink, Johan S.& Koonen, Jos J.. 2019. Inherent Geometry Correction for Diffusion EPI Using the Reference Echoes as Navigators. Concepts in Magnetic Resonance :Part B : Magnetic Resonance Engineering.،Vol. 2019, no. 2019, pp.1-8.
https://search.emarefa.net/detail/BIM-1130814

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

van den Brink, Johan S.& Koonen, Jos J.. Inherent Geometry Correction for Diffusion EPI Using the Reference Echoes as Navigators. Concepts in Magnetic Resonance :Part B : Magnetic Resonance Engineering. No. 2019 (2019), pp.1-8.
https://search.emarefa.net/detail/BIM-1130814

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

van den Brink, Johan S.& Koonen, Jos J.. Inherent Geometry Correction for Diffusion EPI Using the Reference Echoes as Navigators. Concepts in Magnetic Resonance :Part B : Magnetic Resonance Engineering.. 2019. Vol. 2019, no. 2019, pp.1-8.
https://search.emarefa.net/detail/BIM-1130814

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1130814