Validity and Reliability of the Newly Developed Surface Electromyography Device for Measuring Muscle Activity during Voluntary Isometric Contraction

المؤلفون المشاركون

Shin, Myung Jun
Jang, Myung Hun
Ahn, Se Jin
Lee, Jun Woo
Rhee, Min-Hyung
Chae, Dasom
Kim, Jinmi

المصدر

Computational and Mathematical Methods in Medicine

العدد

المجلد 2018، العدد 2018 (31 ديسمبر/كانون الأول 2018)، ص ص. 1-9، 9ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2018-08-29

دولة النشر

مصر

عدد الصفحات

9

التخصصات الرئيسية

الطب البشري

الملخص EN

Objective.

The purpose of this study was to establish the validity and reliability of the newly developed surface electromyography (sEMG) device (PSL-EMG-Tr1) compared with a conventional sEMG device (BTS-FREEEMG1000).

Methods.

In total, 20 healthy participants (10 males, age 30.3 ± 2.9 years; 10 females, age 22.3 ± 2.7 years) were recruited.

EMG signals were recorded simultaneously on two devices during three different isometric contractions (maximal voluntary isometric contraction (MVIC, 40% MVIC, 80% MVIC)).

Two trials were performed, and the same session was repeated after 1 week.

EMG amplitude recorded from the dominant biceps brachii (BB) and rectus femoris (RF) muscles was analyzed for reliability using intrasession intraclass correlation coefficient (ICC).

Concurrent validity of the two devices was determined using Pearson’s correlation coefficient.

Results.

Nonnormalized sEMG data showed moderate to very high reliability for all three contraction levels (ICC = 0.832–0.937 (BB); ICC = 0.814–0.957 (RF)).

Normalized sEMG values showed no to high reliability (ICC = 0.030–0.831 (BB); ICC = 0.547–0.828 (RF)).

sEMG signals recorded by the PSL-EMG-Tr1 showed good to excellent validity compared with the BTS-FREEEMG1000, at 40% MVIC (r = 0.943 (BB), r = 0.940 (RF)) and 80% MVIC (r = 0.983 (BB); r = 0.763 (RF)).

Conclusions.

The PSL-EMG-Tr1 was performed with acceptable validity.

Furthermore, the high accessibility and portability of the device are useful in adjusting the type and intensity of exercise.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Jang, Myung Hun& Ahn, Se Jin& Lee, Jun Woo& Rhee, Min-Hyung& Chae, Dasom& Kim, Jinmi…[et al.]. 2018. Validity and Reliability of the Newly Developed Surface Electromyography Device for Measuring Muscle Activity during Voluntary Isometric Contraction. Computational and Mathematical Methods in Medicine،Vol. 2018, no. 2018, pp.1-9.
https://search.emarefa.net/detail/BIM-1131974

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Jang, Myung Hun…[et al.]. Validity and Reliability of the Newly Developed Surface Electromyography Device for Measuring Muscle Activity during Voluntary Isometric Contraction. Computational and Mathematical Methods in Medicine No. 2018 (2018), pp.1-9.
https://search.emarefa.net/detail/BIM-1131974

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Jang, Myung Hun& Ahn, Se Jin& Lee, Jun Woo& Rhee, Min-Hyung& Chae, Dasom& Kim, Jinmi…[et al.]. Validity and Reliability of the Newly Developed Surface Electromyography Device for Measuring Muscle Activity during Voluntary Isometric Contraction. Computational and Mathematical Methods in Medicine. 2018. Vol. 2018, no. 2018, pp.1-9.
https://search.emarefa.net/detail/BIM-1131974

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1131974