Comparative Analysis of Bone Tissue Temperature during Implant Preparation with Variable Drilling Parameters: In Vitro Study
المؤلفون المشاركون
Pelc, Anna
Mikulewicz, Marcin
Kosior, Piotr
المصدر
العدد
المجلد 2020، العدد 2020 (31 ديسمبر/كانون الأول 2020)، ص ص. 1-9، 9ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2020-03-19
دولة النشر
مصر
عدد الصفحات
9
التخصصات الرئيسية
الملخص EN
Purpose.
The aim of this work was to compare the temperature fluctuations that occur during the development of the implant bed using three different implant systems and the impact on their value of cooling method and rotational speed of drill.
Material and Methods.
As a model of the human jaw due to the analogy of bone structure and hardness, pig ribs were used.
Drills from three different implant systems were used in the study: Straumann® (Straumann GmbH, Basel, Switzerland), AnyRidge® (Megagen Implant Co., Ltd., Daegu, South Korea), and Osstem (OSSTEM IMPLANT CO., LTD., Seoul, South Korea).
The sequence of three successive drills was given—from pilot drill to final drill.
For each system, a group with two water cooling methods, without cooling, and three different speed ranges, 800, 1200, and 1500 rpm, and their effect on temperature fluctuations was evaluated.
The temperature was measured by thermography.
Results.
The highest temperature increases were noted during preparation with pilot drills.
The maximum temperature (50.8°C) was noted for the AnyRidge pilot drill at 1500 rpm without cooling.
When cooling with physiological saline, none of the applied drills exceeded 28°C.
Significant differences between lack of cooling and cooling with saline at 20°C and 3°C have been demonstrated.
During preparation with cooling, the difference between the times of the maximum temperature achievement was observed between AnyRidge® and Osstem (2.6 vs.
1.6 s, p=0.004).
Conclusion.
The experiment showed that the drills of the tested implant systems differed in the amount of heat generated during operation.
The temperature of the cooling solution and the rotational speed applied have an influence on its amount.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Kosior, Piotr& Pelc, Anna& Mikulewicz, Marcin. 2020. Comparative Analysis of Bone Tissue Temperature during Implant Preparation with Variable Drilling Parameters: In Vitro Study. BioMed Research International،Vol. 2020, no. 2020, pp.1-9.
https://search.emarefa.net/detail/BIM-1136854
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Kosior, Piotr…[et al.]. Comparative Analysis of Bone Tissue Temperature during Implant Preparation with Variable Drilling Parameters: In Vitro Study. BioMed Research International No. 2020 (2020), pp.1-9.
https://search.emarefa.net/detail/BIM-1136854
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Kosior, Piotr& Pelc, Anna& Mikulewicz, Marcin. Comparative Analysis of Bone Tissue Temperature during Implant Preparation with Variable Drilling Parameters: In Vitro Study. BioMed Research International. 2020. Vol. 2020, no. 2020, pp.1-9.
https://search.emarefa.net/detail/BIM-1136854
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-1136854
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر