In Vitro Evaluation of Manual Torque Values Applied to Implant-Abutment Complex by Different Clinicians and Abutment Screw Loosening

المؤلفون المشاركون

Kose, T. Emre
Karatasli, Burcin
Erdem, Mehmet Ali
Demircan, Sabit
Dincer Kose, Onur
Cene, Erhan
Aya, Serhan Aydın
Cankaya, Abdulkadir Burak

المصدر

BioMed Research International

العدد

المجلد 2017، العدد 2017 (31 ديسمبر/كانون الأول 2017)، ص ص. 1-9، 9ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2017-04-03

دولة النشر

مصر

عدد الصفحات

9

التخصصات الرئيسية

الطب البشري

الملخص EN

Preload is applied to screws manually or using a torque wrench in dental implant systems, and the preload applied must be appropriate for the purpose.

The aim of this study was to assess screw loosening and bending/torsional moments applied by clinicians of various specialties following application of manual tightening torque to combinations of implants and abutments.

Ten-millimeter implants of 3.7 and 4.1 mm diameters and standard or solid abutments were used.

Each group contained five implant-abutment combinations.

The control and experimental groups comprised 20 and 160 specimens, respectively.

Implants in the experimental group were tightened by dentists of different specialties.

Torsional and bending moments during tightening were measured using a strain gauge.

Control group and implants with preload values close to the ideal preload were subjected to a dynamic loading test at 150 N, 15 Hz, and 85,000 cycles.

The implants that deformed in this test were examined using an optical microscope to assess deformities.

Manual tightening did not yield the manufacturer-recommended preload values.

Dynamic loading testing suggested early screw loosening/fracture in samples with insufficient preload.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Dincer Kose, Onur& Karatasli, Burcin& Demircan, Sabit& Kose, T. Emre& Cene, Erhan& Aya, Serhan Aydın…[et al.]. 2017. In Vitro Evaluation of Manual Torque Values Applied to Implant-Abutment Complex by Different Clinicians and Abutment Screw Loosening. BioMed Research International،Vol. 2017, no. 2017, pp.1-9.
https://search.emarefa.net/detail/BIM-1138549

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Dincer Kose, Onur…[et al.]. In Vitro Evaluation of Manual Torque Values Applied to Implant-Abutment Complex by Different Clinicians and Abutment Screw Loosening. BioMed Research International No. 2017 (2017), pp.1-9.
https://search.emarefa.net/detail/BIM-1138549

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Dincer Kose, Onur& Karatasli, Burcin& Demircan, Sabit& Kose, T. Emre& Cene, Erhan& Aya, Serhan Aydın…[et al.]. In Vitro Evaluation of Manual Torque Values Applied to Implant-Abutment Complex by Different Clinicians and Abutment Screw Loosening. BioMed Research International. 2017. Vol. 2017, no. 2017, pp.1-9.
https://search.emarefa.net/detail/BIM-1138549

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1138549