Electromagnetic Interference from Left Ventricular Assist Device (LVAD)‎ Inhibiting the Pacing Function of an Implantable Cardioverter-Defibrillator (ICD)‎ Device

المؤلفون المشاركون

Erqou, Sebhat
Kormos, Robert L.
Wang, Norman C.
McNamara, Dennis M.
Bazaz, Raveen

المصدر

Case Reports in Cardiology

العدد

المجلد 2018، العدد 2018 (31 ديسمبر/كانون الأول 2018)، ص ص. 1-4، 4ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2018-10-03

دولة النشر

مصر

عدد الصفحات

4

التخصصات الرئيسية

الأمراض

الملخص EN

There is an increasing prevalence of patients with concomitant implantable cardioverter-defibrillators (ICDs) and left ventricular devices (LVADs).

The potential for negative interactions between these continually evolving technologies is a valid concern.

Previously reported interactions include inappropriate ICD therapy and interference with ICD telemetry function.

Understanding the nature of such interactions and developing a comprehensive strategy to approach such situations are important.

In this report, we describe a case of electromagnetic interference from LVAD inhibiting the pacing function of an ICD that was corrected by reprograming the device.

We would encourage investigators to review patients with ICD and LVAD in their institutions in order to help assess the frequency and nature of these and other interactions.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Erqou, Sebhat& Kormos, Robert L.& Wang, Norman C.& McNamara, Dennis M.& Bazaz, Raveen. 2018. Electromagnetic Interference from Left Ventricular Assist Device (LVAD) Inhibiting the Pacing Function of an Implantable Cardioverter-Defibrillator (ICD) Device. Case Reports in Cardiology،Vol. 2018, no. 2018, pp.1-4.
https://search.emarefa.net/detail/BIM-1141112

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Erqou, Sebhat…[et al.]. Electromagnetic Interference from Left Ventricular Assist Device (LVAD) Inhibiting the Pacing Function of an Implantable Cardioverter-Defibrillator (ICD) Device. Case Reports in Cardiology No. 2018 (2018), pp.1-4.
https://search.emarefa.net/detail/BIM-1141112

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Erqou, Sebhat& Kormos, Robert L.& Wang, Norman C.& McNamara, Dennis M.& Bazaz, Raveen. Electromagnetic Interference from Left Ventricular Assist Device (LVAD) Inhibiting the Pacing Function of an Implantable Cardioverter-Defibrillator (ICD) Device. Case Reports in Cardiology. 2018. Vol. 2018, no. 2018, pp.1-4.
https://search.emarefa.net/detail/BIM-1141112

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1141112