Fracture of the Femoral Component Caused by Insufficient Crimping after Modern Unicompartmental Knee Arthroplasty

المؤلفون المشاركون

Yamamoto, Tetsuji
Ishibashi, Yoichi
Mashiba, Tasuku
Mori, Masaki

المصدر

Case Reports in Orthopedics

العدد

المجلد 2019، العدد 2019 (31 ديسمبر/كانون الأول 2019)، ص ص. 1-5، 5ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2019-12-03

دولة النشر

مصر

عدد الصفحات

5

التخصصات الرئيسية

الطب البشري

الملخص EN

Fracture of a femoral component after modern unicompartmental knee arthroplasty is very rare.

Although this is not the first case on this subject, no study has reported insufficient crimping as the cause of femoral component loosening that led to breakage of a metallic component.

A 69-year-old man underwent medial unicompartmental knee arthroplasty for right medial knee osteoarthritis.

His early postoperative course was good; however, the 1-year postoperative radiograph showed an apparent radiolucent line around the femoral component, and he occasionally had right knee pain.

However, he had been followed up conservatively because he had been doing well even while doing heavy agricultural work.

At 8 years after surgery, because breakage of the femoral component was found, revision surgery was performed using bicruciate-retaining total knee arthroplasty.

The removed fractured femoral component revealed a thick cement mantle detached from the bone surface.

The postoperative course of the patient after the revision surgery was excellent.

We suggest that the causes of femoral component breakage include early implant loosening caused by uneven cement crimping of the femoral component to the bone and excessive loading stress as a result of heavy labour.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Ishibashi, Yoichi& Mashiba, Tasuku& Mori, Masaki& Yamamoto, Tetsuji. 2019. Fracture of the Femoral Component Caused by Insufficient Crimping after Modern Unicompartmental Knee Arthroplasty. Case Reports in Orthopedics،Vol. 2019, no. 2019, pp.1-5.
https://search.emarefa.net/detail/BIM-1142715

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Ishibashi, Yoichi…[et al.]. Fracture of the Femoral Component Caused by Insufficient Crimping after Modern Unicompartmental Knee Arthroplasty. Case Reports in Orthopedics No. 2019 (2019), pp.1-5.
https://search.emarefa.net/detail/BIM-1142715

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Ishibashi, Yoichi& Mashiba, Tasuku& Mori, Masaki& Yamamoto, Tetsuji. Fracture of the Femoral Component Caused by Insufficient Crimping after Modern Unicompartmental Knee Arthroplasty. Case Reports in Orthopedics. 2019. Vol. 2019, no. 2019, pp.1-5.
https://search.emarefa.net/detail/BIM-1142715

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1142715