Tulip-Screw Head Disjunction from Posterior C2 Fracture Fixation Instrumentation
المؤلفون المشاركون
Burley, Halle E. K.
Ansari, Darius S.
von Glinski, Alexander
Goodmanson, Ryan
Schell, Benjamin
Chapman, Jens R.
Oskouian, Rod J.
المصدر
العدد
المجلد 2020، العدد 2020 (31 ديسمبر/كانون الأول 2020)، ص ص. 1-5، 5ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2020-02-24
دولة النشر
مصر
عدد الصفحات
5
التخصصات الرئيسية
الملخص EN
This report presents an unusual case of instrumentation failure after posterior fixation of a C2 fracture and reviews currently available treatment alternatives.
The patient, a 53-year-old female, initially presented to the emergency department at an outside facility with acute alcohol intoxication and acute neck pain following a fall from a ladder.
CT demonstrated bilateral C2 pars fractures and unstable posteroinferior displacement of the posterior elements.
She underwent an emergent C2 open-reduction internal fixation (ORIF) at the outside facility with 3.5 mm polyaxial synapse pedicle screws (DePuy Synthes, Switzerland).
There were no known complications and the patient was discharged.
Two years after the index operation, cervical CT scan at a different facility revealed that although the fracture was fully healed, bilateral tulip caps had detached from the pedicle screw heads at C2.
All implants were removed without postoperative complications.
Industry review of alternate lag screws approved for the cervical spine demonstrated that there is not currently an ideal implant for fixation of C2 fractures without fusion.
Cannulated trauma screws, which are low profile and would have avoided the instrumentation failure seen here, are not currently FDA approved for the cervical spine.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Burley, Halle E. K.& Ansari, Darius S.& von Glinski, Alexander& Goodmanson, Ryan& Schell, Benjamin& Chapman, Jens R.…[et al.]. 2020. Tulip-Screw Head Disjunction from Posterior C2 Fracture Fixation Instrumentation. Case Reports in Orthopedics،Vol. 2020, no. 2020, pp.1-5.
https://search.emarefa.net/detail/BIM-1149939
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Burley, Halle E. K.…[et al.]. Tulip-Screw Head Disjunction from Posterior C2 Fracture Fixation Instrumentation. Case Reports in Orthopedics No. 2020 (2020), pp.1-5.
https://search.emarefa.net/detail/BIM-1149939
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Burley, Halle E. K.& Ansari, Darius S.& von Glinski, Alexander& Goodmanson, Ryan& Schell, Benjamin& Chapman, Jens R.…[et al.]. Tulip-Screw Head Disjunction from Posterior C2 Fracture Fixation Instrumentation. Case Reports in Orthopedics. 2020. Vol. 2020, no. 2020, pp.1-5.
https://search.emarefa.net/detail/BIM-1149939
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-1149939
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي