Comparison of Low-Pass Filters for SPECT Imaging

المؤلفون المشاركون

Sayed, Inayatullah S.
Ismail, Siti S.

المصدر

International Journal of Biomedical Imaging

العدد

المجلد 2020، العدد 2020 (31 ديسمبر/كانون الأول 2020)، ص ص. 1-9، 9ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2020-04-01

دولة النشر

مصر

عدد الصفحات

9

التخصصات الرئيسية

الطب البشري

الملخص EN

In single photon emission computed tomography (SPECT) imaging, the choice of a suitable filter and its parameters for noise reduction purposes is a big challenge.

Adverse effects on image quality arise if an improper filter is selected.

Filtered back projection (FBP) is the most popular technique for image reconstruction in SPECT.

With this technique, different types of reconstruction filters are used, such as the Butterworth and the Hamming.

In this study, the effects on the quality of reconstructed images of the Butterworth filter were compared with the ones of the Hamming filter.

A Philips ADAC forte gamma camera was used.

A low-energy, high-resolution collimator was installed on the gamma camera.

SPECT data were acquired by scanning a phantom with an insert composed of hot and cold regions.

A Technetium-99m radioactive solution was homogenously mixed into the phantom.

Furthermore, a symmetrical energy window (20%) centered at 140 keV was adjusted.

Images were reconstructed by the FBP method.

Various cutoff frequency values, namely, 0.35, 0.40, 0.45, and 0.50 cycles/cm, were selected for both filters, whereas for the Butterworth filter, the order was set at 7.

Images of hot and cold regions were analyzed in terms of detectability, contrast, and signal-to-noise ratio (SNR).

The findings of our study indicate that the Butterworth filter was able to expose more hot and cold regions in reconstructed images.

In addition, higher contrast values were recorded, as compared to the Hamming filter.

However, with the Butterworth filter, the decrease in SNR for both types of regions with the increase in cutoff frequency as compared to the Hamming filter was obtained.

Overall, the Butterworth filter under investigation provided superior results than the Hamming filter.

Effects of both filters on the quality of hot and cold region images varied with the change in cutoff frequency.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Sayed, Inayatullah S.& Ismail, Siti S.. 2020. Comparison of Low-Pass Filters for SPECT Imaging. International Journal of Biomedical Imaging،Vol. 2020, no. 2020, pp.1-9.
https://search.emarefa.net/detail/BIM-1169171

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Sayed, Inayatullah S.& Ismail, Siti S.. Comparison of Low-Pass Filters for SPECT Imaging. International Journal of Biomedical Imaging No. 2020 (2020), pp.1-9.
https://search.emarefa.net/detail/BIM-1169171

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Sayed, Inayatullah S.& Ismail, Siti S.. Comparison of Low-Pass Filters for SPECT Imaging. International Journal of Biomedical Imaging. 2020. Vol. 2020, no. 2020, pp.1-9.
https://search.emarefa.net/detail/BIM-1169171

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1169171