الملخص EN

Human blood is an important medical detection index.

With the development in clinical medical detection instruments and detection technology, the requirements for detection accuracy and efficiency have been gradually improved.

Fluorescent immunochromatography is a new detection technique.

It has the characteristics of high efficiency, convenience, no pollution, and wide detection range.

Human blood can be detected quickly using fluorescent immunochromatography.

At present, it has received great attention from the field of clinical testing.

In this paper, a set of fluorescent immunochromatographic analyzer has been designed.

It is mainly based on the principle of fluorescence immunochromatography.

A new method of signal analysis and system design for fluorescent immunochromatography analyzer is proposed.

By using the improved threshold function denoising algorithm, the quantitative detection of fluorescent immunochromatographic strip is realized.

The concentration of pathogenic factors (cancer cells) in human serum can be measured conveniently and accurately.

The system integrates many peripheral modules, including fluorescence signal acquisition, fluorescence signal processing, quantitative curve fitting, and test results.

In this paper, the quantitative detection experiments of the system are carried out in three aspects: linearity, repeatability, and sensitivity.

The experimental results show that the linear correlation coefficient is up to 0.9976, and the limit of detection is up to 0.05 ng/ml.

The requirements of the system are satisfied.

The system performance is good, and the quantitative result is accurate.

Therefore, the establishment of a fluorescence analysis system is of great significance.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Shao, Lei& Zhang, Longyu& Li, Shilin& Zhang, Pengyuan. 2019. Design and Quantitative Analysis of Cancer Detection System Based on Fluorescence Immune Analysis. Journal of Healthcare Engineering،Vol. 2019, no. 2019, pp.1-9.
https://search.emarefa.net/detail/BIM-1175059

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Shao, Lei…[et al.]. Design and Quantitative Analysis of Cancer Detection System Based on Fluorescence Immune Analysis. Journal of Healthcare Engineering No. 2019 (2019), pp.1-9.
https://search.emarefa.net/detail/BIM-1175059

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Shao, Lei& Zhang, Longyu& Li, Shilin& Zhang, Pengyuan. Design and Quantitative Analysis of Cancer Detection System Based on Fluorescence Immune Analysis. Journal of Healthcare Engineering. 2019. Vol. 2019, no. 2019, pp.1-9.
https://search.emarefa.net/detail/BIM-1175059

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1175059